Quick test
DTA 100 C
48 / 90
822-129-8
A quick test is used to make a quick evaluation of the oil status. In this process, an
orientation value will be determined for the breakdown strength of the insulating oil.
Quick test parameters:
Standard: none;
Stirrer: inactive;
Output voltage: max. possible output voltage.
It is not possible to adjust the quick test parameters. To adjust individual parameters, create
a user-defined measurement. Further information: Chapter
User-defined measurement
(on
page 51)
Main menu
>
Quick test
1. In the main menu, select the menu item
Quick test
.
2. Select the menu item
Next
to confirm the selection.
A prompt appears on the display to set the electrode gap. You can choose any electrode
shape you want during a quick test.
Recommendation:
Select an electrode shape based on the selected standard.
10
Q
UICK TEST