PC PROGRAMM / REGISTER „POWER“
29/32
MB112EN - 11185046, 19A1, Baumer_HENQ1100_OI_EN
Field 6:
”Pin Levels”
This is the check on signal levels�
If the high and low values are achieved, it is comfirmed by a tick/check mark in the check
box�
K#: Identified signal
K#+high: HIGH level (HTL ≥ 80%), (TTL ≥ 2.5 V)
K#+low: LOW level (HTL ≤ 20%), (TTL ≤ 1.0 V)
K#-high: Inverted signal HIGH level (HTL ≥ 80%), (TTL ≥ 2.5 V)
K#-low: Inverted signal LOW level (HTL ≤ 20%), (TTL ≤ 1.0 V)
8.5
Register „Power“
The power is calculated from the measured supply voltage to the encoder and the mea-
sured current that is drawn�
Field 1
Field 2
Fig.20:
Register “Power“
Field 1: ”HTL [V]”
The measured voltage (in volts) at the supply pins 9+10 of the SUB-D socket, see
, which normally supplies HTL encoders, but also (regulated) TTL
encoders�
Field 2: ”TTL [V]”
The measured voltage (in volts) at the supply pins 11+12 of the SUB-D socket, see
, which normally supplies TTL encoders and sinewave encoders�