SETUP
MB112EN - 11185046, 19A1, Baumer_HENQ1100_OI_EN 8/32
Display 6:
Position of the reference signal (
Z=282°
)
This value gives the position (in degrees) of the reference signal relative to the position
of the encoder after the supply voltage for the HENQ 1100 is switched on or the encoder
changed�
5.9
Display measurements as graphics (menu
)
A bar graph display is used for online presentation of the measured values�
The values indicated are, see also
Column 1:
The speed value ( shown at left, next to plain text with mathematical sign)�
The scaling is made so that the highest bar corresponds to the value for the menu item
MaxSpeed
set in menu
�
Column 3 and 4:
This value gives the position and the position of the reference signal relative to the
position of the encoder after the supply voltage for the HENQ 1100 is switched on or the
encoder changed�
The lowest bar corresponds to 0° and the highest corresponds to 360°�
Column 6 and 7:
The pulse/pause ratio of the encoder signals K1 and/or K2, whereby the middle of the
bar corresponds to a 50:50 ratio� The highest or lowest value, respectively, refers to the
maximum error
MS_Error
as set up in menu
(for example 30:70 resp� 70:30)�
Column 9:
The phase shift of K1 relative to the rising edge of K2, whereby the middle of the bar
corresponds to 90°�
In menu
the maximum permitted phase shift can be set up in the menu item
PhaseErr
�
The lowest bar corresponds to a [90° -
PhaseErr
] and the highest corresponds to a
[90° +
PhaseErr
]�
If a reference signal is available, its first appearance is acknowledged by
K0=OK!
�
If the reference signal fails then even once, this will be detected and signaled by
K0=ERR
�