Operating Instructions
Conductivity measurement
CombiLyz AFI4/AFI5
Inductive conductivity transmitter
5.3 Mounting and dismounting the AFI5 cable
Mounting the AFI5 cable
f
To plug the cable in, align the 2 red points.
A locking mechanism ensures that the cable
cannot be pulled out unintentionally.
Dismounting the AFI5 cable
f
Press both ends of the locking mechanism
together.
f
Pull the cable out.
5.4 Mounting the AFIx
WARNING
Danger to health from contaminated medium
f
Only use weld-in sleeves or adapters from Baumer.
f
Do not seal the process connections with Teflon tape (PTFE) or elastomer.
f
Welding work must only be carried out by welders trained in the area of hygiene.
9
Hole for mounting the sensor is easily accessible and dry.
9
Vessel is drained.
9
Mounting location and sensor orientation meet the conditions in chapter “5.1 Mounting conditions” on
f
Mount weld-in sleeve or adapter as follows:
3-A mark or arrow points upwards
Leakage hole points downwards
Hygienically and internally flush
f
Grind welding to Ra ≤ 0.8.
f
Screw in sensor.
Tightening torque: 20 … 25 Nm
f
Check leak-tightness of the sleeve.
f
Check leak-tightness of the cable glands or M12 plug.
f
Ensure that the housing cover is fastened properly.
www.baumer.com
7 / 52