Operating Instructions
Conductivity measurement
CombiLyz AFI4/AFI5
Inductive conductivity transmitter
4. Transport and storage
f
Check packaging and sensor for damage.
f
In the event of damage: Do not use sensor.
f
Store sensor where it will be secure against shock.
Storage temperature: –30 … 80 °C
Relative humidity: < 98 %
5. Mounting
5.1 Mounting conditions
Fig. 3. Mounting locations
The sensor can be mounted at any point on the
vessel or pipe.
Fig. 4. Fully submerged in medium
For correct operation the sensor must be fully
submerged in the medium.
1
Fig. 5.
Mounting in flow direction
To allow a sufficient self-cleaning to happen, the
hole through the sensor must be oriented in the
flow direction. Arrows (1) on the sensor above the
connection mark the orientation of the hole.
≥1 m
Fig. 6. Recommended distance from bend
To avoid problems due to turbulence in flowing
media, Baumer recommends to mount the sensor
in a distance of at least 1 m from a bend.
www.baumer.com