Operating Instructions
Conductivity measurement
CombiLyz AFI4/AFI5
Inductive conductivity transmitter
f
Select conductivity range.
Menu
Input config.
Conductivity ranges
Conductivity range 1
Conductivity range 2
Conductivity range 3
Conductivity range 4
Select range
Select range
Select range
Select range
AFIx menu
f
If desired, select or define further settings:
Concentration output
Display colors
Warnings
Relays
8.2
Configuration via FlexProgram
9
FlexProgrammer 9701 is connected.
Configuration options with the AFIx transmitter:
Select HART or IO-Link mode
Configure switch output
Configure temperature compensation
Select temperature source for compensation
Select conductivity range
Set current limits
Select concentration output
Configure media screen
Make a data logging
Calibrate sensor and media
Configuration options with the DFON display:
Select screen layout
Select backlight intensity
Define relays
Define warning and error indication
Calibrating the sensor (conductivity/temperature)
9
AFIx is powered up
9
The used medium has a known conductivity/temperature
f
Adjust an offset for the conductivity/temperature for the sensor.
f
If necessary, reset the sensor offset in FlexProgram.
Calibrating a medium
For the different ranges it is possible to calibrate the medium by calculating the temperature compensation
after 3 measurements.
9
Calibration is done in a controlled environment
9
AFIx is powered up
f
Have the sensor tip immersed in the medium for 1 minute before measuring the conductivity.
f
Measure the conductivity at 3 different temperatures using the same medium.
f
Calculate the temperature compensation for the medium.
For further information, refer to the HELP section in FlexProgram.