DIP SWITCHES
The LC-4C has two eight-position dip switches for setting certain gain ranges and defaults.
Switches A to E set the default gain range of the working electrode. Dip switch F is not
used. The default gain range is automatically set at power-up, and will be in effect until a
change is made with the RANGE control. The factory-set default gain range is 0.5
µ
Afs. You
may want to reset this default to the gain at which you typically work. Then the LC-4C will
automatically re-equilibrate after a power outage. Dip switch positions are as follows:
Power-up
Dip switch position
gain
A
B
C
D
E
0.1 nAfs
DN
DN
DN
DN
DN
0.2 nAfs
UP
DN
DN
DN
DN
0.5 nAfs
DN
UP
DN
DN
DN
1.0 nAfs
UP
UP
DN
DN
DN
2.0 nAfs
DN
DN
UP
DN
DN
5.0 nAfs
UP
DN
UP
DN
DN
10 nAfs
DN
UP
UP
DN
DN
20 nAfs
UP
UP
UP
DN
DN
50 nAfs
DN
DN
DN
UP
DN
0.1
µ
Afs
UP
DN
DN
UP
DN
0.2
µ
Afs
DN
UP
DN
UP
DN
0.5
µ
Afs
UP
UP
DN
UP
DN
1.0
µ
Afs
DN
DN
UP
UP
DN
2.0
µ
Afs
UP
DN
UP
UP
DN
5.0
µ
Afs
DN
UP
UP
UP
DN
10
µ
Afs
UP
UP
UP
UP
DN
20
µ
Afs
DN
DN
DN
DN
UP
50
µ
Afs
UP
DN
DN
DN
UP
All other settings of A to E are invalid and will cause the gain display to go blank.
The factory-set default gain range for monitoring the output of the generator electrode (Igen)
is 10
µ
Afs. The default can be reset with the following positions of dip switches G and H:
Gain
Dip switch position
range
G
H
100 nAfs
UP
UP
1.0
µ
Afs
UP
DN
10
µ
Afs
DN
UP
Invalid
DN
DN
The invalid switch setting will cause the decimal point and units to be suppressed on the
Igen
display.
The bottom set of dip switches (marked I to P) allows the user to choose the power-up de-
fault filter setting for the working electrode, the PC remote channel number, and a remote en-
able.
Section 4. Operating the Electrochemical Detector
29