Cause of failure

Solution

Power supply

Check the power supply

 as detailed in Pre-Installation Tests.

Cables

Check the cables

 as detailed in Pre-Installation Tests.

Power Input

If no fault can be found in the power supply or cables, ensure 
the cable length is not too long, causing excessive voltage drop 
between the power supply and magnetometer. Refer to the 
specifications defined in the product brochure.

Magnetometer

In the event of any apparent malfunction, email 

service

@

bartington.com

 or telephone the Bartington Instruments service 

team on +44 (0)1993 706565. Attempted repair or opening of the 
casing by users may invalidate the warranty.

A re-calibration service is available which is traceable to 
international standards.

For information about disposal of the damaged unit, refer to

 

End of 

Life Disposal

.

8. Care and maintenance

8.1. Cleaning the Mag585 and Mag592

Use water and mild soap to remove grime from external enclosures.

Caution:

 Never use chemicals, such as solvents, when cleaning the Mag585 and Mag592.

Caution:

 Take particular care when cleaning around electrical connections. Bent or 

damaged pins may cause the magnetometer to malfunction. 

8.2. Calibration

Return the Mag585 and Mag592 to Bartington Instruments for calibration at the recommended 
intervals. Refer to the Calibration Certificate for further details.

Summary of Contents for Mag585

Page 1: ...Operation Manual for Mag585 Low Radiation Three Axis Magnetic Field Sensor Mag592 Low Radiation Single Axis Magnetic Field Sensor Innovation in Magnetic Measuring Instruments www bartington com...

Page 2: ...ag585 and Mag592 6 5 1 Siting the magnetometer environment recommendations 6 5 2 Connection recommendations 7 5 2 1 Connector pin allocation 7 5 2 2 Interface 7 5 2 3 Power supplies 7 5 3 Pre installa...

Page 3: ...e Mag585 and Mag592 11 8 2 Calibration 11 9 Storage and transport 12 10 End of life disposal 12 10 1 Waste Electrical and Electronic Equipment WEEE Regulations 12 Appendix 1 Use of Alternative Power S...

Page 4: ...e design of these products without prior notice 2 How to use this manual This document describes the installation operation and maintenance of the Mag585 and Mag592 Magnetic Field Sensors Take the tim...

Page 5: ...steresis and low cross talk between axes The Mag592 is a single axis equivalent providing precision measurements of the intensity of static and slowly varying magnetic fields in one direction The Mag5...

Page 6: ...e rock to avoid compromising measurements Note Site the magnetometer several tens of metres from very large ferromagnetic objects that could create fields exceeding the measuring range of the sensor N...

Page 7: ...its Alternatively users may wish to provide their own power supply Refer to Appendix 1 for further information 5 3 Pre installation tests Prior to the installation of the system the magnetometer cable...

Page 8: ...the sensor outputs 5 Confirm that the sum of the measured magnetic field vectors is similar to the expected local earth field 6 Using the Mag585 and Mag592 6 1 Cabling Note The connecting cable to th...

Page 9: ...information may appear due to aliasing This effect is seen in data acquisition systems when the frequency of sampling is lower than the frequency of the signal which is being sampled It may produce a...

Page 10: ...with a 3dB point at 0 1Hz The SCU1 signal conditioning unit provides filters with independent control of the low and high pass filter sections together with offset and gain control for the output of...

Page 11: ...n 44 0 1993 706565 Attempted repair or opening of the casing by users may invalidate the warranty A re calibration service is available which is traceable to international standards For information ab...

Page 12: ...er supplies provide the most suitable methods for connecting to and operating your sensor If you have decided to use an alternative then the information in this Appendix is important Caution Failure t...

Page 13: ...polarity A1 1 Signal power ground The two signal power ground conductors are connected to a common point within the sensor The power supply common power 0V should be connected to only one of them whic...

Page 14: ...cables over 10m The schematic shows signal and power grounds separated at the Host end The Host is fitted with buffer amplifiers configured as Instrumentation Amplifiers IA which have two high impeda...

Page 15: ...the X Y and Z outputs is 0 125V difference to the Host 0V which is the source of power for the sensor However no current flows in the signal ground conductor because the IAs have very high input imped...

Page 16: ...OX28 4GE England T 44 0 1993 706565 E sales bartington com F 44 0 1993 774813 The specifications of the products described in this brochure are subject to change without prior notice Bartington is a...

Reviews: