HELMHOLTZ COIL INSTALLATION AND CALIBRATION
Figure 5: Read Tab - Scaling
•
Orthogonality:
This task requires the use of a separate software. Please contact GMW for details. In addition to the
software, 3 files available from GMW are needed.
The first is a file titled Test Equipment Status.tsf and can be created in notepad if not available. In the
file description below, any entries in () are comments and should not be part of the file
o
Test Equipment Status.tsf
(Start)
[System]
Test Equipment = "HC2" (this may change to HC1, HC9, HC16 depending on the coil used)
System Path = "/C/Bartington/HC2.tef" (the file name should be that of the coil used
–
see above)
Device Path = "/C/Bartington/Mag03_100.tdf" (this file may have a different number based on its
range
–
this file also works for Mag-13)
Aux Path = "/C/Bartington/Mag03_100.tdf" (this file may have a different number based on its
range
–
this file also works for Mag-13)
Callibration Path = "<Not A Path>"
Alarm = 0
[Display Temp]