Bartington Helmholtz Coil Installation And Calibration Download Page 2

 

 

HELMHOLTZ COIL INSTALLATION AND CALIBRATION 

 

 

 

of secondary fields which may affect the signal within the coil. The closer the material is to the coil the 
worse the interferences will become. The presence of conductive material should therefore be carefully 
considered and avoided where possible. For example, the coils should not be set on the ground (where 
metal beams or reinforcing bars may be located), or on or close to metal shelves and tables. 

In  both  cases,  the  size  of  the  object  will  dictate  the  distance  of  clearance  to  the  coils.  In  order  to 

determine whether a site is suitable, a site survey using one or more magnetometer(s) is recommended. 

This can be typically be carried out using Bartington’s Spectramag

-6 and two Mag-13 or Mag690 sensors. 

The  use  of  two  sensors  can  allow  to  perform  gradient  measurements  to  determine  the  differences 
between two points, typically the centre and edge of the homogeneous volume of the coil at the planned 
site. A separate site survey document is also available. 

To  reduce  the  impact  of  local  field  distortion,  the  use  of  the  CU2  closed-loop  module  can  enable 

external  distortion  to  be  cancelled  out  within  the  coil  at  the  location  of  the  reference  magnetometer 
typically mounted in the centre of the coils. However as only one reference sensor is used, the CU2 closed-
loop module does not cancel out the presence of gradients within the coils. Any gradient present will see 
itself being shifted up or down so that the point where the reference magnetometer is located read 0 (or 
whatever value field is applied to the coils). 

Typically, this will mean having a clearance of a few meters (also coil dependent, the larger the coil, 

the larger the clearance area) from structures, shelves, phones, laptop etc… and an increased clearance 

from area accessible by vehicles, especially if these are larger vehicles. 

Any equipment to be used in the vicinity of the coil (tables, c

hairs etc…) should also be checked for 

hygiene  as  for  example  swivelling  chairs  tend  to  have  a  relatively  large  magnetic  signature.  All  these 
elements can be checked either during the site survey stage, or with equipment typically supplied with 
the coils. 

2.

 

System components 

A complete Helmholtz coil system will include the coils, the PA1 Power Amplifier, CU1 Control Unit 

(which  is  used  to  interface  the  PA1  with  an  ADC  card/PC),  the  CU2  Closed-loop  module  (optional  but 
recommended if operating mainly at DC and low frequency below mains power frequencies), a National 
Instruments ADC card and a PC (running the operating software enabling field generation). 

Additionally, a Mag-13MS (range determined by the coil used), Mag-13 cable, Mag-13 Mounting Jig 

will be recommended to be used for initial system calibration, and regular system checks thereafter. 

The clearance recommendations apply not only to elements external to the system, but also apply to 

parts of the system. The coils are supplied with a 3m long cable that connects to the PA1. The whole length 
should be used, and the PA1 should be located as far as possible from the coils, together with the CU1, 
ADC card and PC. 

Summary of Contents for Helmholtz Coil

Page 1: ...e coils the system should be installed in an area clear from permanent fixed ferromagnetic structures such as support beams reinforced concrete or steel shelving as these will distort the local magnet...

Page 2: ...ed up or down so that the point where the reference magnetometer is located read 0 or whatever value field is applied to the coils Typically this will mean having a clearance of a few meters also coil...

Page 3: ...ces 4 Assembly and Orientation The HC1 HC9 and HC16 come pre assembled Only the HC2 requires assembly on site in addition to the 2m Ferronato coil Assembly instructions for the HC2 can be found in the...

Page 4: ...the Mag 13MS mounting jig can be placed within the coils using the necessary support pillars table provided coil dependent The jig should be mounted so that the sensor s Z axis along the length of the...

Page 5: ...MS in the centre of the coil using the provided mounting jig Opening the Bartington Helmholtz Coil Control Software set the coil to the correct type Generator tab settings should be set to o Set the f...

Page 6: ...e field amplitude reading ensure that you are selecting the correct measuring range for the sensor On the Mag 13 the number that follows is the full range of the sensor and corresponds to 10V The aim...

Page 7: ...the offset checks return to the generator tab The settings should be as follow o Set the amplitude to of the sensor range or the maximum field which can be generated by the coils whichever is the lowe...

Page 8: ...into account the sensor s scaling error In RMS this value will be 70 71067 the field applied in the generator tab or the adjusted value to account for the sensor s scaling error o Set the filter to 1...

Page 9: ...tries in are comments and should not be part of the file o Test Equipment Status tsf Start System Test Equipment HC2 this may change to HC1 HC9 HC16 depending on the coil used System Path C Bartington...

Page 10: ...see table at the end of the file Test Equipment HC2 use the same name as the tef file name and the test equipment entry in the test equipment status tsf file DAC Sample Rate Multiplier 4 000000E 0 De...

Page 11: ...tor Z Coil Trim Factor Max coil field HC1 5 000000E 3 5 000000E 3 5 000000E 3 5 000000E 4 HC2 7 330000E 3 6 740000E 3 7 870000E 3 2 000000E 4 HC9 or HC16 1 410000E 3 1 225000E 3 1 050000E 3 1 000000E...

Page 12: ...7E 6 70E 6 100uT full range 10E 6 100E 6 250uT full range 25E 6 250E 6 500uT full range 50E 6 500E 6 1000uT full range 100E 6 1000E 6 Once the files have been created and are stored at C Bartington t...

Page 13: ...The same procedure will apply for Y and Z As the method is however similar for the other two axes we will only provide photos showing the position of the sensor and the potentiometers which need adjus...

Page 14: ...on 0 XYZ Sensor Match XYZ Coil With the sensor in position 0 select the X axis and record the two values given in the alignment section of the software These will be angular errors for Y and Z and wil...

Page 15: ...HELMHOLTZ COIL INSTALLATION AND CALIBRATION Figure 8 Y and Z errors due to X Next rotate the sensor along its X axis so that is in position 1...

Page 16: ...s calculated using the Y X and Z X potentiometers adjust the potentiometers to the target value Once achieved you can return the sensor to position 0 and check that the value obtained has the same abs...

Page 17: ...d Z errors due to Y Xy Zy Rotate the sensor along the Y axis to end up in position 2 see Figure 12 The errors obtained will be Xy and Zy The target value for Xy and Zy will be calculated as Xy Xy 2 an...

Page 18: ...rs due to Z Xz Yz Rotate the sensor along the Z axis to end up in Position 3 see Figure 14 The errors obtained will be Xz and Yz The target value for Xz and Yz will be calculated as Xz Xz 2 and Yz Yz...

Page 19: ...ft on permanently though not necessarily generating field The system can be checked at regular intervals using the supplied equipment Any changes to the local environment should lead to a check of the...

Reviews: