Final inspection 

After production, the micropumps have to pass a final inspection. 

They are tested concerning the maximum flow 

and back pressure.

 

Measurement conditions: 

Pumping media: distilled water 
Temperature: room temperature 23°C 
Controller: extended box mp-x 
Electrical Input: amplitude of 250 Vpp and SRS-Signal with 100 Hz (for liquids) and 300 Hz (for gases) 

Measurements with sensors: 

volume flow 

range: 

0 - 10 ml/min 

accuracy: +/- 1% FS (=0.1 ml/min) 

pressure 

range: 

 1 bar 

accuracy: +/- 0.35% (= 3.5 mbar) 

 

Other application specific outgoing inspections can be offered for all pumps upon customer demand. 

In order to guarantee proper function of the delivered goods and exclude transportation damages please check the 
incoming devices according to specifications after receipt. 
Based on these results a replacement can be carried out within 14 days after delivery free of costs. 

Operating the micropump 

In this chapter the operation of the micropump during the evaluation is described. This chapter provides information 
on the proper connection of the pump with tubing and electrical cables and typical driving parameters to start the 
evaluation. 

 

Please connect suitable tubes to the inlet and outlet. The tubing should have an inner diameter of ~1.3 mm for the 
mp6, mp6-AIR, mp6-AIR-HP and mp6-pi and ~1.2 mm for the mp6-pp. The micropump has to be connected to a 
suitable controller as described in the next sections. 

  

The electrical connection of the micropumps flexible connector is described in the corresponding chapter. The flexible 
connector will be attached to the desired controller. 

On both type pumps, traces of surface corrosion may appear on the electrical connector. This corrosion only causes 
a visual influence, negative effects on the pump performance could not be detected based on performance tests 

Tygon; ID = 1,30 mm; AD = 3 mm 
Tygon; ID = 1,02 mm; AD = 2,72 mm 

 

Tygon; D = 1,30 mm; 

AD = 3 mm 

Tygon; D = 1,02 mm; 

AD = 2,72 mm 

inlet 

flexible connector 
 

outlet 
 

Pumping direction 
 

Summary of Contents for microComponents mp6-gas

Page 1: ...1 Rev 3 1 08 2020 Operating Manual for Micropump mp6 series...

Page 2: ...specifications 5 Technical Data of the mp6 hyb1 5 Technical Data of the mp6 liq1 2 6 Technical specifications mp6 pi1 7 Technical specifications mp6 pp1 9 Technical specifications mp6 gas1 10 Final in...

Page 3: ...Low Voltage Directive 2014 35 EU Description of functions The micropumps have been developed for the transport of gases or liquids The controllers have been developed for operating the micropumps Bar...

Page 4: ...ditions including vapors is not permitted Staff selection and qualification All work in connection with the installation assembly commissioning decommissioning disassembly operation servicing cleaning...

Page 5: ...in typ flow rate p 0 8 ml min 4 typ back pressure p 0 500 mbar 7 25 psi 4 Gases air typ volume flow p 0 20 ml min 5 typ back pressure p 0 80 mbar 1 16 psi 5 1 Typical values Values can vary under appl...

Page 6: ...Please contact us for more information Please find more information concerning the controller and the equipment in the corresponding manuals mp6 liq Order code mp6 liq Pump type piezoelectric diaphrag...

Page 7: ...1 1811 x 0 5906 x 0 1498 in Weight 2 g Fluidic connectors barbed tube clip outer diameter 1 9 mm length 3 5 mm 2 Electric connector flex connector 1 25 mm pitch Power consumption 50 mW 5 Self priming...

Page 8: ...Tygon tubing 1 3 mm inner diameter 3 Conditions Suction pressure 10 mbar DI water settings mp x 100 Hz 250 V SRS the max volume flow will be reached after a few minutes of operation time 4 Conditions...

Page 9: ...ter Controllable flow range8 8 l min 4000 l min typ flow rate p 0 4 ml min 4 typ back pressure p 0 500 mbar 7 25 psi 4 1 Typical values Values can vary under application conditions Content is subject...

Page 10: ...1498 in Weight 2 g Fluidic connectors barbed tube clip outer diameter 1 9 mm length 3 5 mm 2 Electric connector flex connector 1 25 mm pitch Power consumption 150 mW 5 Self priming yes 3 Pumping medi...

Page 11: ...er 3 Conditions air room temperature 23 C settings mp x 300 Hz 250 V SRS the max volume flow will be reached after a few minutes of operation time 4 Conditions gases room temperature 23 C mp x 300 Hz...

Page 12: ...rating the micropump In this chapter the operation of the micropump during the evaluation is described This chapter provides information on the proper connection of the pump with tubing and electrical...

Page 13: ...mp6 series to the cable The pumps can be connected via a FCC connector The layout of the connector and pin assignment of the pumps is as shown below Each piezo P1 P2 has a single lead for the negative...

Page 14: ...n top or bottom of the pump the Molex connector is not water tight additional sealing with e g silicone will be necessary fixation of the Molex connector in the final assembly is recommended The Molex...

Page 15: ...neral for driving the pump with air the following points should be taken into account Higher frequencies should be used compared to pumping water Large amplitudes respectively driving voltages should...

Page 16: ...uencies The performance depends on the environmental conditions Combination of micropumps The micropumps can be combined to achieve higher flow rates or pressure than the single unit can generate With...

Page 17: ...l 49 231 47730 500 Fax 49 231 47730 501 Visit our Website www bartels mikrotechnik de for further information on applications Tutorials and helpful answers to frequently asked questions can be found i...

Reviews: