P-600 SIC Analyzer Manual
Chapter VI: Maintenance & Service
42
4. Bleed off existing tank pressure. This is done via the valve on the right side of the tank
assemblies (can vary) when facing the rear of the unit (see below).
5. If you have a filling station installed on your system, skip to step 6. Otherwise remove the plugs at
the top of the tank and carefully pour your solvent into the tank. Use the color-coded liquid level
indicators on the tank’s front face as a guide. Repeat for other tanks as necessary. The tanks
hold roughly 10 gallons (37.8 liters). Replace the plugs when finished and skip to step 10.
6. Make sure the air supply to the solvent filling station is on (see below).
7. Connect a flexible tube (Teflon or braided SS) from the filling station pump (pumps are labeled
per solvent -
1/2” Swagelok connection) to your solvent holding vessel (see below).
8. Open the supply valve (positions vary) (see below).