P-600 SIC Analyzer Manual
Chapter IV: Programming
27
System Settings
The System Settings submenu allows you to establish global operating parameters for the Analyzer.
Use the Index key to move to the desired menu item. Use the Up/Down Arrow keys to change the displayed
setting or value. Tap Enter key to reset cycle counter or reload defaults.
General
Those are general setting for Analyzer’s operation.
•
Conc. Unit
—
This menu selection allows you to select the unit of measure the Analyzer will use
when displaying and outputting measurement data. The choices are PTB, mg/L and g/m
³
.
•
Temp. Unit
—
This menu selection allows you to select the temperature scale. The choices are
°C or °F.
•
Stdby Mode
—
this controls how the Analyzer will operate upon application of power. When OFF
is selected, the instrument will begin performing analysis automatically when power is applied.
When ON is selected, the instrument will power up in the Offline mode and must be manually
placed Online.
•
Come Read
—
This menu item allows you to set the length of time the Analyzer’s “come read”
relay will be activated upon the completion of a measurement cycle.
•
Num Cycles
—
This is the number of measurements which have been performed since the cycle
counter was last reset. To reset the counter, touch the Enter key.
•
Max Cycles
—
This menu item allows the operator to set the number measurements which can
be performed before the Analyzer’s Maintenance Alarm is activated.