24
drA:
Adaptive tune enable.
drA.t:
Adaptive tune trigger level in display
units. Range = 1 9999.
Pid LiSt
G.SP (Gain Setpoint):
Is the temperature at
which the controller switches from the (SEt1)
PID values to the (SEt 2) PID values.
Pb:
Proportional band in display units. (SEt 1)
ti:
Integral time in seconds. (SEt 1)
td:
Derivative time in seconds. (SEt 1)
Pb2:
Proportional band. (SEt 2)
ti2:
Integral time in seconds. (SEt 2)
td2:
Derivative time in seconds. (SEt 2)
ACCS LiSt (Access List)
Access Code (Code needed to enter or
change the other configuration parameters
which are not normally accessible.) Not
accessable.
Alarms
The controller will flash an alarm message in
the home display if an alarm condition is
detected.
IdHi:
PV deviation high alarm.
2FSH:
PV full scale high alarm.
8 S
EGMENT
& 4
X
16 S
EGMENT
P
ROGRAMMABLE
M
ODELS W
/OTP
Note
The following alarm messages are facto-
ry default settings and may vary if you
have changed the configuration of your
controller:
IDHi: =
50°C
2FSH = 1225°C