23
rmP.U:
Press the UP or DOWN ARROW to
toggle between ramp units (SEc, min or
Hour). Controller will default to “SEc.”
dwL.U:
Press the UP or DOWN ARROW to
toggle between dwell units (SEc, min or
Hour). Controller will default to “SEc.”
Cyc.n:
Press the UP or DOWN ARROW to
set the number of program cycles (1 to 999 or
cont). The controller will default to “cont.”
SEG.n:
Press the UP or DOWN ARROW to
select the segment number (1-8 in 8 segment
models, 1-16 in 4x16 models).
tYPE:
Press the UP or DOWN ARROW to
select the segment type [End (end of pro-
gram), rmP.r = ramp rate (ramp to a specified
setpoint at a set rate), rmp.t = ramp time
(ramp to a specified temperature in a set
time), dwEll (to maintain a constant tempera-
ture for a set time), StEP (climb instanta-
neously from current to specified tempera-
ture), cALL (to call a program as a subroutine,
available only on 4x16 programmable mod-
els)]. The controller will default to “End.” Other
parameters used with tYPE include; tGt target
setpoint), Rate (rate of temperature increase)
and dur (time to target setpoint or time to
dwell).
End.t:
End segment type: dwELL (dwell con-
tinuous), rSEt (reset) and S OP (End
Segment Output power level.
AL LiSt (Alarm List)
IdHi:
Deviation High Alarm.
Atun LiSt: (Autotune List)
tunE:
One-shot autotune enable.
8 S
EGMENT
& 4
X
16 S
EGMENT
P
ROGRAMMABLE
M
ODELS W
/OTP