Dimensions and Clearance Requirements
Dimensions
13.5” W x 19.5” H x 17.0” D (34.3 x 49.5 x 43.2 cm)
Clearances
Sides - 9” (22.9 cm) minimum for servicing.
Above - 3” (7.6 cm) minimum for removal of the top cover.
Front - 4.75” (12.1 cm) minimum for opening the front door.
Feed Water Requirements
Types
RO, DI, distilled.
TOC
< 1.0 ppm.
Turbidity
1.0 N.T.U. maximum.
Pressure Range
Gravity feed to 100 psig (7kg/cm2) maximum.
Temperature Range
4°C - 40°C (40-104°F)
TDS (CaCO3)
< 70 ppm
Silica
< 1 ppm
Silt
< 5% SDI
Product Water
Water Quality
Resistivity
> 18.0 M
Ω
-cm
TOC
< 2.0 ppb
Flow Rate
1.5 lpm maximum at minimum inlet feed water pressure 10 PSIG with a new final filter.
Volumetric Dispense
Accuracy: ± 5%
Repeatability: ± 3%
TOC Analyzer
Range
1 to 250 ppb
Accuracy
±1 ppb or ±15% (whichever is greater)
Cycle Time
< 3.5 minutes
Water Consumption
< 15 ml per cycle
Min. Water Resistivity
> 10 M
Ω
for specified accuracy