14
Barnstead
STANDBY
START
STOP
DISPENSE
NANOpure DIamond
NANOpure DIamond
“
BACK
ENTER
NANOpure DIamond TOC - UV Control Panel
Main Power Switch
The main power switch on the NANOpure DIamond
TOC - UV is located on the lower back right side of the
unit (as you face the front of the unit), directly above the
power cord receptacle.
Control Panel
The NANOpure DIamond TOC - UV is controlled through
a panel which incorporates switches to control its func-
tions: START/STOP, STANDBY, DISPENSE, DOWN, UP,
ENTER and BACK. This control panel utilizes a digital
display to show system information and the purity of the
product water. The intensity (brightness) of the display
may be adjusted according to user preference. See
“Adjusting Display Brightness” in the “User Settings” sec-
tion of this manual.
A single LED located on the top of the control panel will
illuminate to inform you of the system status as follows:
•
LED is not illuminated: System is OFF or in
(Idle) mode.
•
Solid Green LED: System is recirculating and
operating normally.
•
Blinking Green LED: System is in Standby
mode.
•
Red LED: System is alerting user to an
operational error or maintenance issue.
The control panel can be removed from the unit and
remotely mounted. Please refer to “Installing the Control
Panel in a Remote Location” in the “Installation” section of
this manual for more information about how to remove the
control panel from the unit.
Controls
Note
The removable control panel was not
designed to be repeatedly removed
from the unit, therefore, you may
experience difficulty when attempting
to remove it. It is therefore recom-
mended that the control panel be per-
manently mounted in a remote location
or remain intact with the unit.
LED