44
6.
Make sure the UV lamp cable is tucked behind
the chamber before closing and relatching the
door.
7.
Reconnect the power cord. Turn unit ON.
8.
Reset UV Timer according to “Resetting the UV
Timer.”
9.
Operate normally.
TOC Lamp Replacement
1.
Place unit in the “Idle” mode.
2.
Turn power off. Disconnect the power cord from
the power entry module.
3.
Open right side door and locate the TOC board
at the bottom. The lamp chamber is protruding
from the board on the right side.
4. Locate the clips on both sides of lamp chamber.
While gently pulling out chamber, wedge a flat
head screwdriver between chamber and clips to
release chamber.
5.
When installing the new lamp chamber, be sure
to carefully insert pins of chamber into left side
row of sockets on TOC board. See Chamber
Replacement photo.
6.
Finish installing the new lamp chamber by
securing with clips. Clips may need to be
pushed apart so the chamber will snap back into
place.
7.
Reconnect power cord and reset the TOC lamp
timer. (See “Resetting the TOC Lamp Timer.”)
M
AINTENANCE AND
S
ERVICING
Chamber
Pins
Sockets
used for Pin
Insertion
Chamber Replacement
Clips