Normal Unit Operation
Below is a brief description of the normal automatic
operation of the EASYpure RO. If the unit does not
operate as described below, refer to the Trouble-
shooting Guide.
When you activate the power switch, the follow-
ing events will occur:
1.
All LEDs light momentarily to indicate that
they work.
2.
After approximately 5 seconds, the auto-
matic solenoid valves are energized. The
inlet valve opens, supplying the EASYpure
RO with water. The flush valve also
opens, diverting virtually all the water to
drain.
3.
After approximately 20 seconds, the pump
energizes. The majority of the water pro-
duced goes to drain.
4.
10 seconds after the pump energizes, the
pressure switch begins monitoring the
incoming water pressure.
5.
After 3 minutes, the flush solenoid valve
closes and the EASYpure RO begins mak-
ing product water and sending it to the
storage reservoir.
6.
The unit begins to monitor membrane per-
formance and the storage reservoir floats
begin to monitor tank levels 5 seconds
after the flush solenoid valve closes.
During operation, the storage reservoir
floats will: 1) switch the EASYpure RO into
standby mode (pump off) when the stor-
age reservoir is full, 2) switch the
EASYpure RO back into water production
mode (pump on) when the reservoir level
Service and Repair
Warning
Disconnect from the power
supply prior to maintenance
and servicing.
Refer servicing to qualified
personnel.
Ensure all piping connections
are tight to avoid leakage of
chemicals.
Ensure adequate ventilation
when using chemicals for
cleaning.
Follow carefully the manufac-
turers’ safety instructions on
labels of chemical containers
and Material Safety Data
Sheets (M.S.D.S.).