15
Connections to Equipment
Supplied By the EASYpure
RO.
Your EASYpure RO also includes two
lengths of tubing with quick disconnect
inserts attached. One is 1/4" O.D. tubing and
one is 3/8" O.D. tubing. Use the
1/
4" tubing
to feed directly from the EASYpure RO to
another EASYpure unit. Use the 3/8" O.D.
tubing to feed the accessory 30 Liter
Reservoir if one is being used.
1.
Install the quick disconnect insert
attached to the tubing into one of
the product water quick disconnect
couplings on the EASYpure RO.
(See page 11.)
2.
Install the other end of the tubing
into the unit being fed by the
EASYpure RO according to the
instructions in that unit's manual.
3.
Locate the pump interlock cord
supplied with the EASYpure RO.
Plug one end of the pump interlock
cord into the pump interlock on the
EASYpure RO. (See page 11.)
Remove the jumper plug from the
pump interlock on the other
EASYpure unit and save it for
future use. Plug the other end of
the cord into the pump interlock on
the other EASYpure unit.
I
NSTALLATION
Do not connect the product water
tubing to a connected EASYpure unit
until you have completed the mem-
brane flush procedure for the
EASYpure RO.
Note
The product water quick dis-
connect inserts can be
installed in either product water
quick disconnect coupling. The
quick disconnect ends of the
inserts are the same size.
Note
Note
If you are not now feeding another
EASYpure unit, save the pump
interlock cord for possible future
use.