background image

7

1/8" NPT x 1/4" OD straight connector
provided to connect the Bantam Holder
inlet to the raw water supply. If the raw
water supply service connection is not
3/8" NPT, use another brass fitting as
required.

3.

Cut a piece of tubing from the 2.1m (7')
of 1/4" OD tubing (part no. TU1494X1)
provided to make the plumbing service
connection between the raw water faucet
and the plumbing service        connection
marked “inlet” on the Bantam Holder.
Ensure that sufficient tubing is left to
drain the holder through the plumbing
service connection marked “drain” on the
Bantam Holder to the sink and to make
the plumbing service      connection
between the holder effluent connection
at the top of the unit and the point of
distribution.

4.

Thread the 1/8" NPT x 1/4" OD straight
connector provided (part no. 15451) into
the brass bushing at the faucet opening
and connect the tubing between this
connection and the plumbing service
connection marked “inlet” on the holder.

5.

Ensure that the 1/8" NPT x 1/4" OD
polypropylene elbow (part no. 15847) is
threaded into the effluent connection at
the top of the holder and connect tubing
between this connection and the point of
distribution. See 

Elbow-Tubing Assembly

in this manual to ensure that the
connection is made properly to prevent
leakage.

I

NSTALLATION

Summary of Contents for 494 Series

Page 1: ...1 Type D0800 Bantam Cartridge Holder OPERATION MANUAL AND PARTS LIST SERIES 494 BARNSTEAD THERMOLYNE CORPORATION 60119 3 12 98 Model Voltage D0800 120 D0805 240...

Page 2: ...trapure Mixed Resin 5 Installation 6 Unpacking 6 Choosing a Site 6 Mounting 6 Plumbing Service Connections 6 Elbow Tubing Assembly 8 Cartridge Installation 9 Operation 10 Purity Meter Operation 10 Mai...

Page 3: ...ughout the manual Warnings To avoid electrical shock always 1 Use a properly grounded electrical outlet of correct voltage and current handling capacity 2 Disconnect from the power supply prior to mai...

Page 4: ...i u q e r l a c i r t c e l E 0 0 8 0 D l e d o M 1 z H 0 6 0 5 V 0 2 1 e s a h p 2 z H 0 6 0 5 V 0 2 1 e s a h p 1 2 5 0 8 0 D 1 z H 0 6 0 5 V 0 4 2 e s a h p 2 z H 0 6 0 5 V 0 4 2 e s a h p 1 2 Note...

Page 5: ...the cartridge is new Ultrapure Mixed Resin This cartridge catalog no D0809 has the capacity to remove 65 grams 750 grains of dissolved ionized impurities as NaCl About 3 4 75 of the capacity is capabl...

Page 6: ...Mounting The Bantam Holder is shipped assembled for bench mounting Four screws are provided for wall mounting if desired Plumbing Service Connections The Bantam Holders are supplied with polypropylen...

Page 7: ...drain on the Bantam Holder to the sink and to make the plumbing service connection between the holder effluent connection at the top of the unit and the point of distribution 4 Thread the 1 8 NPT x 1...

Page 8: ...nd backup ring in the adapter nut and install the nut on the elbow Do not install the o ring at this time 4 Push the tubing through the nut until it bottoms out in the elbow 5 Remove the adapter nut t...

Page 9: ...wn to allow the cartridge to be centered between the top and bottom o ring 3 Turn the lower tightening wheel until the cartridge seats against the top o ring INSTALLATION Note Water purified by the io...

Page 10: ...ing Meter by setting the on off switch to the on position When the purity meter drops below the acceptable minimum purity or when the purity meter needle registers in the 0 05 to 0 025 megohm red rang...

Page 11: ...er needle is aligned with the line indicating the beginning of the meter scale The zero adjustment screw is located on the face of the meter below the scales 3 To calibrate the meter further connect a...

Page 12: ...12 Exploded View...

Page 13: ...x D O 4 1 s s a r b e v l a V 7 V 5 1 1 8 3 5 1 2 V 0 3 2 9 1 0 3 3 r e t e M 8 7 9 5 5 1 r e d l o h e g d i r t r a C 9 6 9 8 5 1 e l p p i N 0 1 2 0 5 5 1 l e e h W g n i n e t h g i T r e w o L 1...

Page 14: ...whom you purchased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle your re...

Page 15: ...15...

Page 16: ...ng and contamination must be returned to our factory and if upon inspection it is concluded that failure is not due to excessive high temperature or contamination warranty replacement will be provided...

Reviews: