32
Nio
R
EADING
ROOM
DEFINITION
SUBMENU
*
*This menu is available only in MDNC-2121, MDNC-3121, MDNG-2121
and MDNG-6121.
C
ALIBRATION
I
NFORMATION
SUBMENU
*
*This menu is available only in MDNC-2121, MDNC-3121, MDNG-2121
and MDNG-6121.
Reading room
Indicates the reading room type you are
editing. You can select another room to
edit by rotating the control wheel.
Max. ambient light
This indicates the maximum ambient light
that corresponds to the selected reading
room.
If the measured ambient light is higher
than the value entered here, you should
take measures to darken the room or
select another reading room type.
Preset corr. value
This value is taken into account in the
calculation of the DICOM DF if DICOM Offset
is set to “Preset”.
To each reading room type corresponds
another preset value.
(Preset) Ambient
value
Shows the ambient light correction value
taken into account to calculate the DICOM
DF, expressed in lux.
Bright luminance
Shows the bright luminance value taken
into account to calculate the DICOM DF.
Dark luminance
Shows the dark luminance value taken into
account to calculate the DICOM DF.
Ambient correction
Shows the ambient light correction value
taken into account to calculate the DICOM
DF, expressed in cd/m².