K5902105 (451920611795) /04
MDSC-2232
57
Mission critical applications
We strongly recommend there is a replacement monitor immediately available in mission critical applications.
Use of Electrical Surgical Knives
Provide as much distance as possible between the electrosurgical generator and other electronic equipment
(such as monitors). An activated electrosurgical generator may cause interference with them. The interference
can activate the OSD menu of the display and as such disrupt the functionality of the display.
Power connection
–
Equipment with external 24 VDC power supply
•
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC (
) SELV power supply.
•
The medical approved DC (
) power supply must be powered by the AC mains voltage.
•
The power supply is specified as a part of the ME equipment or combination is specified as a ME system.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Transient over-voltage
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
•
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd. Ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
•
To maintain compliance with EMC Regulation, use only shielded interface cables for the connection to
peripheral devices.
Connection of PEMS by network/data coupling to other equipment (MDSC-2232 MNA
version only)
•
Connection of the PEMS to a network/data coupling that includes other equipment could result in
previously unidentified risks to patients, operators or third parties.
•
The responsible organization should identify, analyze, evaluate and control these risks.
•
Subsequent changes to the network/data coupling could introduce new risks and require additional
analysis.
•
Changes to the network/data coupling include:
-
changes in network/data coupling configuration;
-
connection of additional items to the network/data coupling configuration;
-
disconnecting items from the network/data coupling configuration;
-
update of equipment connected to the network/data coupling configuration;
-
upgrade of equipment connected to the network/data coupling configuration
Power cords:
•
Europe: H05VV-F or H05VVH2-F PVC cord with appropriate EU plug.
US and Canada:
“
hospital grade
”
cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent receptacle
marked hospital only or hospital grade. These instructions need to be marked either on the equip. or on a
tag on the power cord
•
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
•
Mains lead protection: Power cords should be routed so that they are not likely to be walked upon or
pinched by items placed upon or against them, paying particular attention to cords at plugs and
receptacles.
•
The power supply cord should be replaced by the designated operator only at all time.
•
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Summary of Contents for MDSC-2232 Series
Page 6: ...K5902105 451920611795 04 MDSC 2232 6...
Page 7: ...7 K5902105 451920611795 04 MDSC 2232 Welcome 1...
Page 10: ...K5902105 451920611795 04 MDSC 2232 10 Welcome...
Page 11: ...11 K5902105 451920611795 04 MDSC 2232 Parts controls and connectors 2...
Page 20: ...K5902105 451920611795 04 MDSC 2232 20 Parts controls and connectors...
Page 21: ...21 K5902105 451920611795 04 MDSC 2232 Display installation 3...
Page 28: ...K5902105 451920611795 04 MDSC 2232 28 Display installation...
Page 29: ...29 K5902105 451920611795 04 MDSC 2232 Daily operation 4...
Page 38: ...K5902105 451920611795 04 MDSC 2232 38 Daily operation...
Page 39: ...39 K5902105 451920611795 04 MDSC 2232 Advanced operation 5...
Page 54: ...K5902105 451920611795 04 MDSC 2232 54 Advanced operation...
Page 55: ...55 K5902105 451920611795 04 MDSC 2232 Important information 6...