Immunity test
IEC 60601-1-2 4
th
edition
(2014)
Test levels
Compliance level
Electromagnetic
environment
–
guidance
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user of
the MDSC-2232 requires
continued operation
during power mains
interruptions, it is
recommended that the
MDSC-2232 be powered
from an uninterruptible
power supply or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V/m (150 kHz to 80
MHz)
9 to 28 V/m in
communication service
channels up to 6 GHz
3 V/m (150 kHz to 80
MHz)
9 to 28 V/m in
communication service
channels up to 6 GHz
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
MDSC-2232, including
cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800
MHz
d = 2.3
√
P
800 MHz to 2.5
Ghz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the recommended
separation distance in
meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey,
2
should be less
than the compliance level
in each frequency range.
3
1:
is the a.c. mains voltage prior to application of the test level.
2:
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the MDSC-2232 is used exceeds the applicable RF compliance level above, the MDSC-2232
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the MDSC-2232.
3:
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Summary of Contents for MDSC-2232 Series
Page 6: ...K5902105 451920611795 04 MDSC 2232 6...
Page 7: ...7 K5902105 451920611795 04 MDSC 2232 Welcome 1...
Page 10: ...K5902105 451920611795 04 MDSC 2232 10 Welcome...
Page 11: ...11 K5902105 451920611795 04 MDSC 2232 Parts controls and connectors 2...
Page 20: ...K5902105 451920611795 04 MDSC 2232 20 Parts controls and connectors...
Page 21: ...21 K5902105 451920611795 04 MDSC 2232 Display installation 3...
Page 28: ...K5902105 451920611795 04 MDSC 2232 28 Display installation...
Page 29: ...29 K5902105 451920611795 04 MDSC 2232 Daily operation 4...
Page 38: ...K5902105 451920611795 04 MDSC 2232 38 Daily operation...
Page 39: ...39 K5902105 451920611795 04 MDSC 2232 Advanced operation 5...
Page 54: ...K5902105 451920611795 04 MDSC 2232 54 Advanced operation...
Page 55: ...55 K5902105 451920611795 04 MDSC 2232 Important information 6...