43
K5902142 /04
Nio 5MP
6.5 EMC notice
General information
This device is for use in professional healthcare facility environments only.
With the installation of the device, use only the delivered external cables and power supply or a spare part
provided by the legal manufacturer. Using another can result in a decrease of the immunity level of the device.
WARNING:
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
WARNING:
Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Nio 5MP,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
Electromagnetic emissions
The Nio 5MP is intended for use in the electromagnetic environment specified below. The customer or the
user of the Nio 5MP should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment –
Guidance
RF emissions
CISPR 11
Group 1
The Nio 5MP uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Nio 5MP is suitable for use in
all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
This Nio 5MP complies with appropriate medical EMC standards on emissions to, and interference from
surrounding equipment. Operation is subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Nio 5MP is intended for use in the electromagnetic environment specified below. The customer or the
user of the Nio 5MP should assure that it is used in such an environment.
Summary of Contents for MDNG-6221
Page 1: ...ENABLING BRIGHT OUTCOMES MDNG 6221 option NC EM User Guide 5 8MP high bright display Nio 5MP...
Page 5: ...5 K5902142 04 Nio 5MP Welcome 1...
Page 8: ...K5902142 04 Nio 5MP 8 Welcome...
Page 9: ...9 K5902142 04 Nio 5MP Installation 2...
Page 16: ...K5902142 04 Nio 5MP 16 Installation...
Page 17: ...17 K5902142 04 Nio 5MP Daily operation 3...
Page 34: ...K5902142 04 Nio 5MP 34 Advanced operation...
Page 35: ...35 K5902142 04 Nio 5MP Cleaning your display 5...
Page 37: ...37 K5902142 04 Nio 5MP Important information 6...
Page 52: ...K5902142 04 Nio 5MP 52 Important information...
Page 53: ......
Page 54: ...www barco com K5902142 04 2022 01 10 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...