27
K5902142 /04
Nio 5MP
3.
Enter the
Color Presets
submenu.
4.
Select one of the available Color Presets and confirm.
4.14 Ambient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display when the display function
is set to DICOM. Please refer to
.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light correction
value into account. This value is determined by the selected reading room. Therefore, it is also important to
select a realistic reading room when enabling ALC. This can be done by following the instructions in
To enable/disable ALC:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Ambient Light Compensation
submenu.
4.
Select
Enabled/Disabled
as desired and confirm.
4.15 Reading rooms
About reading rooms
Reading rooms can only be selected on your display when the display function is set to DICOM.
Please refer to
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
“Ambient Light Compensation (ALC)”, page 27
to enable ALC.
The available reading rooms for your Nio 5MP are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Office:
Corresponds to light conditions in office rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading room:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Reading Room
submenu.
Summary of Contents for MDNG-6221
Page 1: ...ENABLING BRIGHT OUTCOMES MDNG 6221 option NC EM User Guide 5 8MP high bright display Nio 5MP...
Page 5: ...5 K5902142 04 Nio 5MP Welcome 1...
Page 8: ...K5902142 04 Nio 5MP 8 Welcome...
Page 9: ...9 K5902142 04 Nio 5MP Installation 2...
Page 16: ...K5902142 04 Nio 5MP 16 Installation...
Page 17: ...17 K5902142 04 Nio 5MP Daily operation 3...
Page 34: ...K5902142 04 Nio 5MP 34 Advanced operation...
Page 35: ...35 K5902142 04 Nio 5MP Cleaning your display 5...
Page 37: ...37 K5902142 04 Nio 5MP Important information 6...
Page 52: ...K5902142 04 Nio 5MP 52 Important information...
Page 53: ......
Page 54: ...www barco com K5902142 04 2022 01 10 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...