5. Advanced operation
5.11 I-Luminate default mode
About the I-Luminate default mode
The I-Luminate mode de
fi
nes the default activated mode during the I-Luminate boost. This mode can be:
Screen
The boost mode is applied on the screen display.
Film (large or small)
Film (large or small): the boost mode activates a
high illuminated rectangle on top of the screen,
simulating a light box for use with radiological
fi
lm.
The size of this rectangle is similar to a classic
(large or small) radiological
fi
lm.
To set the I-Luminate mode:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > I-Luminate
menu.
3. Enter the
Default mode
submenu.
4. Select
Screen/Small Film/Large Film
as desired and con
fi
rm.
5.12 I-Luminate
fi
lm position
About the I-Luminate
fi
lm position
By default, the I-Luminate rectangle comes up in the top/center of the screen. This position can be changed
to top/left or top/right corner.
To set the I-Luminate
fi
lm position:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > I-Luminate
menu.
3. Enter the
Film Position
submenu.
4. Select
Left/Center/Right
as desired and con
fi
rm.
5.13 Luminance target
About the luminance target
The luminance target of your Mammo Tomosynthesis is adjustable over a prede
fi
ned range. When you
change the luminance target, the display will adjust its backlight to reach the target.
To set the luminance target:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration
menu.
3. Enter the
Luminance Target
submenu.
4. Set a luminance target value as desired and con
fi
rm.
The default, factory DICOM calibrated luminance value is available in the technical spec-
i
fi
cations table. The guaranteed backlight lifetime is valid for this setting.
K5902035
MAMMO
TOMOSYNTHESIS
07/12/2017
27