Immunity test
IEC 60601-1-2 4
th
edition
(2014)
Test levels
Compliance level
Electromagnetic
environment
–
guidance
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
continued operation
during power mains
interruptions, it is
recommended that the
MDSC-8255 be powered
from an uninterruptible
power supply or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.