1. Safety

1.5

HD for fully enclosed projection systems

HD

Hazard Distance (HD) is the distance measured from the projection lens at which the intensity or the energy per surface
unit becomes lower than the applicable exposure limit on the cornea or on the skin. The light beam is considered (to
be) unsafe for exposure if the distance from a person to the light source is less than the HD.

Restriction Zone (RZ) based on the HD

The projector is also suitable for rear projection applications; projecting a beam onto a defuse coated projection screen. As displayed
in image 1-3 two areas should be considered: the restricted enclosed projection area (RA) and the observation area (TH).

RA

TH

sw

PD

HD

DIFFUSE

sw

RZ

sw

sw

PR

HD

REFLECTION

RESTRICTED

AREA

RESTRICTED

AREA

Image 1-3
RA Restricted Access location (enclosed projection area).
PR Projector.
TH Theater (observation area).
RZ Restriction Zone.
PD Projection Distance.
SW Separation Width. Must be minimum 1 meter.

For this type of setup 3 different HD shall be considered:

HD as discussed in "High Brightness precautions: Hazard Distance (HD)", page 6 , relevant for intrabeam exposure.

HD

re

fl

ection

: the distance that has to be kept restrictive related to the re

fl

ected light from the rear projection screen.

HD

diffuse

: the relevant distance to be considered while observing the defuse surface of the rear projection screen.

As described in "High Brightness precautions: Hazard Distance (HD)", page 6 , it is mandatory to create a restricted zone within
the beam areas closer than any NOHD. In the enclosed projection area the combination of two restricted zones are relevant: The
restricted zone of the projected beam toward the screen; taking into account 1 meter Separation Width (SW) from the beam onward.
Combined with the restricted zone related to the rear re

fl

ection from the screen (HD

re

fl

ection

); also taking into account a 1 meter lateral

separation.

The HD

re

fl

ection

distance equals 25% of the difference between the determined HD distance and the projection distance to the rear

projection screen. To determine the HD distance for the used lens and projector model see graphs in chapter "HD in function of
modifying optics", page 9 
.

HD

reflection

= 25% (HD – PD)

The light emitted from the screen within the observation shall never exceed the RG2 exposure limit, determined at 10 cm. The
HD

diffuse

can be neglected if the measured light at the screen surface is below 5000 cd/m² or 15000 LUX.

8

R5906714 DP2K CLP SERIES 09/05/2016

Summary of Contents for DP2K CLP series

Page 1: ...DP2K CLP series Safety manual R5906714 01 09 05 2016 ...

Page 2: ...Barco NV President Kennedypark 35 8500 Kortrijk Belgium Phone 32 56 36 82 11 Fax 32 56 36 883 86 Support www barco com en support Visit us at the web www barco com Printed in Belgium ...

Page 3: ...cisely All local installation codes should be adhered to Notice on safety This equipment is built in accordance with the requirements of the international safety standards IEC60950 1 EN60950 1 UL60950 1 and CAN CSA C22 2 No 60950 1 which are the safety standards of information technology equipment including electrical business equipment These safety standards impose important requirements on the u...

Page 4: ... responsible for the location Why a restricted access location This is a RG3 product Based on international requirements no person in allowed to enter the projected beam within the zone between the projection lens and the related Hazard Distance HD This shall be physically impossible by creating sufficient separation height or by placing optional barriers Within the restricted area operator traini...

Page 5: ...rojector always take it to an authorized trained service person when service or repair work is required Do not use an accessory attachment which is not recommended by the manufacturer Lightning For added protection for this video product during a lightning storm or when it is left unattended and unused for long periods of time unplug it from the wall outlet and disconnect media and communication c...

Page 6: ...e projector eg during service always disconnect the power from the projector mains terminals and UPS INLET connector Use the disconnect device in the building installation to disconnect the power to the projector mains terminals and unplug the power cord at the UPS INLET connector if the UPS INLET is separately powered Do not place this equipment on an unstable cart stand or table The product may ...

Page 7: ...ot permitted to use the projector in another position See installation procedure for correct installation Special care for Laser Beams Special care should be used when DLP projectors are used in the same room as high power laser equipment Direct or indirect hitting of a laser beam on to the lens can severely damage the Digital Mirror DevicesTM in which case there is a loss of warranty Never place ...

Page 8: ...ns installed See next chapter HD in function of modifying optics page 9 To protect untrained end users as cinema visitors the installation shall comply with the following installation requirements Oper ators shall control access to the beam within the hazard distance or install the product at the height that will prevent spectators eyes from being in the hazard distance Radiation levels in excess ...

Page 9: ...s defined as 1 0m 3 3 ft by the FDA CDRH RA TH PR RZ HD SW 1m SW SW SW HD SH RA TH RZ SH B TOP VIEW A SIDE VIEW Image 1 1 A Side view B Top view RA Restricted Access location boot area of projector PR Projector TH Theater RZ Restriction Zone in the theater SH Separation Height SW Separation Width Based on national requirements no person is allowed to enter the projected beam within the zone betwee...

Page 10: ...flection the distance that has to be kept restrictive related to the reflected light from the rear projection screen HDdiffuse the relevant distance to be considered while observing the defuse surface of the rear projection screen As described in High Brightness precautions Hazard Distance HD page 6 it is mandatory to create a restricted zone within the beam areas closer than any NOHD In the enclo...

Page 11: ...1 Safety 1 6 HD in function of modifying optics DP2K 15CLP Hazard Distance Image 1 4 DP2K 20CLP Hazard Distance Image 1 5 R5906714 DP2K CLP SERIES 09 05 2016 9 ...

Reviews: