Display information
33
• The reading room parameters are pre-defined in the display
according to the AAPM (American Association of Physicists in
Medicine). However they can be changed within certain
limits.
Display function selection
To select a display function (DF):
1.
Open the main menu.
2.
Open the
Display Function
menu.
3.
Select
Display Function
.
4.
Select the desired DF.
5.
Return to the Display Function menu.
You can select the following DFs:
Name
Description
DICOM
Select a DICOM display function for most
medical viewing applications.
The DICOM function results in more visi-
ble grayscales in the images.
Dynamic Gamma 2.2
Dynamic Gamma 1.8
These are gamma functions that are
shifted to take into account the non-zero
luminance of an LCD panel when driven
with a “black” signal.
They are especially useful in CT applica-
tions to improve the perception of low
Haunsfield values.
Native
If you select Native, the native panel
behavior will not be corrected.
Test
This is identical to the Native DF.