background image

29

R5911667 /03

AMM215WTTP

WARNING:

Use of this equipment adjacent to or stacked with other equipment should be avoided

because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.

WARNING:

Use of accessories, transducers and cables other than those specified or provided by

the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.

Electromagnetic emissions

The AMM215WTTP is intended for use in the electromagnetic environment specified below. The customer or
the user of the AMM215WTTP should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment

Guidance

RF emissions
CISPR 11

Group 1

The AMM215WTTP uses RF
energy only for its internal function.
Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby
electronic equipment.

RF emissions
CISPR 11

Class B

The AMM215WTTP is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for
domestic purposes.

Harmonic emissions
IEC 61000-3-2

Class D

Voltage fluctuations/ flicker
emissions
IEC 61000-3-3

Complies

This AMM215WTTP complies with appropriate medical EMC standards on emissions to, and interference
from surrounding equipment. Operation is subject to the following two conditions: (1) this device may not
cause harmful interference, and (2) this device must accept any interference received, including interference
that may cause undesired operation.

Interference can be determined by turning the equipment off and on.

If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:

Reorient or relocate the receiving antenna or equipment.

Increase the separation between the equipment and receiver.

Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.

Consult the dealer or an experienced technician for help.

Electromagnetic immunity

The AMM215WTTP is intended for use in the electromagnetic environment specified below. The customer or
the user of the AMM215WTTP should assure that it is used in such an environment.

Immunity test

IEC 60601 test levels

Compliance level

Electromagnetic
environment

guidance

Electrostatic discharge
(ESD)
IEC 61000-4-2

± 8 kV contact

± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air

Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%

Electrical fast transient /
burst
IEC 61000-4-4

± 2 kV

100 kHz repetition frequency

Mains power quality
should be that of a typical
commercial or hospital
environment

Surge
IEC 61000-4-5

± 0.5 kV, ± 1 kV, ± 2kV

Mains power quality
should be that of a typical

Important Information

Summary of Contents for AMM215WTTP

Page 1: ...ENABLING BRIGHT OUTCOMES User Guide LCD Color Display AMM215WTTP...

Page 2: ...red office Barco NV President Kennedypark 35 8500 Kortrijk Belgium www barco com en support www barco com Barco Inc 3059 Premiere Parkway Suite 400 Duluth GA 30097 USA www barco com en support www bar...

Page 3: ...vanced 16 4 3 Touch Screen Driver Installation 17 4 4 AMM215WTTP Driver Installation Instructions 17 5 Cleaning your display 19 5 1 Cleaning instructions 20 6 Troubleshooting 21 6 1 Troubleshooting li...

Page 4: ...R5911667 03 AMM215WTTP 4...

Page 5: ...gs cautions notes and tips There are four levels of precautionary or advisory statements that may be used in this user guide In descending order of importance they are WARNING Describes hazards or dan...

Page 6: ...Quick start guide Manual Multilingual documentation disc Cable cover screw AC Adaptor part number BPM050S12F09 Bridgepower Your system provider may offer alternative cords or cables depending on the i...

Page 7: ...monitor mounting devices such as wall and table arms To mount the monitor on a UL certified swing arm or other mounting fixture follow the instructions included with the mounting fixture to be used P...

Page 8: ...R5911667 03 AMM215WTTP 8 Welcome...

Page 9: ...n this user guide WARNING Sufficient expertise is required to install this equipment All devices and complete setup must be tested before taking into operation CAUTION When the display is assembled in...

Page 10: ...that could cause a fire or shock hazard Immediately report any shipping damage to the carrier or transportation company and contact customer service in case of return 2 3 Start Your Installation To S...

Page 11: ...11 R5911667 03 AMM215WTTP Image 2 1 Installation...

Page 12: ...Connect the USB cable from the USB port of your display to the USB port of your computer 5 Configure the touch screen Refer to the Touch Screen Driver Installation section of this user guide 6 Once t...

Page 13: ...13 R5911667 03 AMM215WTTP Daily Operation 3...

Page 14: ...nu 2 Press and hold the Menu key with the OSD menu on the screen The OSD menu will disappear Press the Power key once while the Menu key is still being pressed The Lock Unlock menu will appear for 3 s...

Page 15: ...15 R5911667 03 AMM215WTTP Advanced Operation 4...

Page 16: ...ue restored DVI VGA select input source OSD Language OSD menu language selection English French Deutsch Italian Spanish Japanese Traditional Chinese and Simplified Chinese RGB color temperature select...

Page 17: ...Instructions PC Running VISTA XP 2000 If you are using a PC running VISTA XP 2000 1 Power on the PC 2 Be sure the USB cable is connected from the PC to the LCD display 3 Open the CD ROM 4 Follow the s...

Page 18: ...R5911667 03 AMM215WTTP 18 2 Be sure the USB cable is connected to the computer 3 Follow the step by step instructions found in the folder file on the CD ROM Advanced Operation...

Page 19: ...19 R5911667 03 AMM215WTTP Cleaning your display 5...

Page 20: ...ean the monitor with non abrasive cloths and cleaning solutions used in hospitals to clean similar equipment CAUTION Do not apply excessive pressure on the LCD CAUTION To clean the display do not spra...

Page 21: ...21 R5911667 03 AMM215WTTP Troubleshooting 6...

Page 22: ...refresh rate of your graphics card to make it compatible with the LCD display Vague image characters and graphics Use Horizontal size to make an adjustment If the problem still exists use Fine to mak...

Page 23: ...23 R5911667 03 AMM215WTTP Important Information 7...

Page 24: ...the air flow and cause temperature increase of the electronics General recommendations Keep the monitor clean to prolong its operational lifetime LCD panel performance may deteriorate in the long term...

Page 25: ...ling or connecting the camera 9 Do not stack more than 6 boxes high 10 The device has no means to be incorporated in an IT network in the clinical environment 11 The enclosure has to be checked upon c...

Page 26: ...ice or your municipal waste disposal service For details please visit the Barco website at http www barco com AboutBarco weee Turkey RoHS compliance T rkiye Cumhuriyeti AEEE Y netmeli ine Uygundur Rep...

Page 27: ...rmity has been filed for this product A sample of this document may be found in the addendum which accompanies this manual For a copy of the Declaration of Conformity document contact Barco Inc and re...

Page 28: ...s no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and...

Page 29: ...cal EMC standards on emissions to and interference from surrounding equipment Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device mu...

Page 30: ...s short interruptions and voltage variations on power supply input lines IEC 61000 4 11 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 q Mains power quality should be that of a typical commercial o...

Page 31: ...compliance level in each frequency range b Interferency may occur in the vicinity of equipment marked with the following symbol At 80 MHz and 800 MHz the higher frequency range applies These guidelin...

Page 32: ...ation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to th...

Page 33: ...t must be recycled according to the European WEEE Waste Electrical and Electronic Equipment directive Indicates Direct Current DC DC power control switch Fragile handle with care Keep dry This way up...

Page 34: ...limitations set forth below BARCO will at its option either repair any component of its products that prove defective by reason of improper workmanship or materials or BARCO has the exclusive right to...

Page 35: ...uthorization Buyer will be responsible for packaging preferably original container and shipping the defective goods to BARCO or its ASC shipping charges prepaid BARCO or its ASC will return the out of...

Page 36: ...to change without notice Trademarks All trademarks and registered trademarks are property of their respective owners Copyright notice This document is copyrighted All rights are reserved Neither this...

Page 37: ......

Page 38: ...www barco com R5911667 03 2021 04 14 Barco Inc 3059 Premiere Parkway Suite 400 Duluth GA 30097 USA R5911667 03 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...