R5911667 /03
AMM215WTTP
30
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
commercial or hospital
environment
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0° . 45°, 90°, 135°, 180°, 225°, 270°, and 315°
q)
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the AMM215WTTP
requires continued
operation during power
mains interruptions, it is
recommended that the
AMM215WTTP be
powered from an
uninterruptible power
supply or a battery
0% U
T
; 1 cycle and 70 % U
T
; 25/30 cycles
Single phase: at 0°
0 % U
T
; 250/300 cycle
U
T
is the A.C. mains voltage prior to application of the test level.
Summary of Contents for AMM215WTTP
Page 1: ...ENABLING BRIGHT OUTCOMES User Guide LCD Color Display AMM215WTTP...
Page 4: ...R5911667 03 AMM215WTTP 4...
Page 8: ...R5911667 03 AMM215WTTP 8 Welcome...
Page 11: ...11 R5911667 03 AMM215WTTP Image 2 1 Installation...
Page 13: ...13 R5911667 03 AMM215WTTP Daily Operation 3...
Page 15: ...15 R5911667 03 AMM215WTTP Advanced Operation 4...
Page 19: ...19 R5911667 03 AMM215WTTP Cleaning your display 5...
Page 21: ...21 R5911667 03 AMM215WTTP Troubleshooting 6...
Page 23: ...23 R5911667 03 AMM215WTTP Important Information 7...
Page 37: ......