DOscan10 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Overview 

Thank you for selecting the DOscan10 pocket dissolved oxygen tester. 
This user manual provides a step-by-step guide to help you operate

.

 

the tester, please carefully read the following instructions before use. 
 

Installing the Batteries 

1.  Twist the connector collar counter clockwise, pull the connector

.

 

 

away from the tester. 

 

 

2.  Insert two AAA alkaline batteries into the battery compartment, 
 

note polarity. 

 

 

3.  Push the connector into the tester and twist the connector

 

collar 

 

clockwise until tight. 

 

 

Keypad 

 
Key   

 

Function 

 

   

Press and hold the key to switch the tester on or off 

   

Lock or unlock measurement 

   

Exit the calibration, settings and return to the 

 

   

dissolved oxygen measurement 

 

   

Start calibration 

   

Press and hold the key to enter the setup menu 

   

Select an option 

 

   

Confirm the calibration, settings or displayed option 

 

Display 

 
Icon   

 

Description 

 

When the battery voltage falls below the minimum   
power requirements, the icon automatically   
disappears 

 
 
 

 
MEAS 

Indicates that the tester is in the measurement mode 

CAL 

Indicates that the tester is in the calibration mode 

SETUP 

Indicates that the tester is in the setup mode 

ATC 

Indicates that the temperature compensation is enabled 

 

 

 

Prior to Use 

 

Filling the Electrolyte Solution 

1.1  Take out the dissolved oxygen electrode from the carrying case. 
 

Unscrew the membrane cap from the bottom of the electrode, 

 

rinse the inside and outside with distilled water and blot dry. 

1.2  Fill the membrane cap halfway with electrolyte solution. 
 

 

 
1.3  Screw membrane cap back onto the electrode. Some electrolyte 
 

solution will overflow during this process. 

1.4  Check the electrode, ensure that no air bubbles are trapped in the 
 

electrolyte solution and membrane is not creased or damaged. 

 

 

 

Polarizing the Electrode 

2.1  Insert the electrode connector into the connector socket on tester,   
 

ensure that the connector is fully seated. 

 

 

 

2.2  Press and hold the 

Meas

 key to switch on the tester and wait 10   

 

minutes for the electrode to polarize. 

 

 

Remove air bubbles 

Summary of Contents for DOscan10

Page 1: ...DOscan10 Pocket Dissolved Oxygen Tester USER MANUAL Bante Instruments Inc ...

Page 2: ...the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode ATC Indicates that the temperature compensation is enabled Prior to Use Filling the Electrolyte Solution 1 1 Take out the dissolved oxygen electrode from the carrying ...

Page 3: ...rement endpoint Enable Disable default Auto Power Off If enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Disable default Factory Reset If enabled all of the calibration values and current settings will be deleted or reset to the factory defaults the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode ...

Page 4: ...d the single point calibration in the setup menu we recommend that you perform a 100 saturation calibration in the air saturated water If the 2 points calibration is selected the zero oxygen solution needs to be used During the calibration and measurement the temperature sensor on electrode must be immersed in the sample solution completely the solution should keep 0 3 m s of minimum flow rate to ...

Page 5: ... hold the Cal key to enter the setup menu 2 Press the Enter key until the display shows or 3 Press the Cal key the tester enters the temperature calibration mode 4 Place the electrode into a solution with a known accurate temperature and wait for the measurement to stabilize 5 Press the Cal key to modify the temperature value 6 Press the Enter key to save and press the Meas key to return to the me...

Page 6: ... cap is contaminated or the electrolyte solution is depleted Screen shows Electrode is broken Replace the dissolved oxygen electrode 5 I DOscan10 1 Set the barometric pressure and salinity coefficient in the setup menu 2 Rinse the electrode with distilled water 3 Place the electrode into the sample solution and stir gently Wait for the measurement to stabilize and record the reading During the mea...

Page 7: ...ster is one year from the date of shipment Above warranty does not cover the electrode and electrolyte solution Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mo...