DOscan10   

I

    2 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Setup Menu 

The DOscan10 tester contains 8 menu items in the setup menu, the 
following table describes the functions of each option. 

 
Menu Item 

Option and Description 

 

Salinity Coefficient 

Set the salinity compensation coefficient of sample. 

 

0 to 35 g/L (default 0 g/L) 

 

Pressure Coefficient 

Set the barometric pressure coefficient according to 
the local altitude (refer to page 3). 

 

450 to 850 mmHg (default 760 mmHg) 

 

Calibration Points 

Set the number of calibration points. 

 

1 point (default) 

 

2 points 

 

Measurement Unit 

Set the default measurement unit. 

mg/L 

Milligrams per liter (default) 

ppm 

Parts per million 

Percentage saturation 

 

Degrees Celsius (default) 

 

Degrees Fahrenheit 

 

Temperature Calibration 

Refer to page 4. 

 

Reading ± 10° C 

 

Reading ± 10° F 

 

Auto-Hold 

If enabled, the tester will automatically sense and   
lock the measurement endpoint. 

 

Enable 

 

Disable (default) 

 

Auto-Power Off 

If enabled, the tester will automatically switch off   
if no key is pressed within 8 minutes. 

 

Enable 

 

Disable (default) 

 
 
 

 

 

Factory Reset 

If enabled, all of the calibration values and current   
settings will be deleted or reset to the factory   
defaults, the tester must be recalibrated. 

 

Enable 

 

Disable (default)   

 
Setting the Default Option 

1.  In the measurement mode, press and hold the 

Cal

 key to enter 

 

the setup menu. 

 

 

 

 
2.  If necessary, press the 

Cal

 key again to select an option. 

 

 

 

 
3.  Press the 

Enter

 key, the tester saves the current option and moves 

 

to the next menu item. 

 

 

 

 
4.  Repeat steps above until the tester returns to the measurement 
 

mode. 

 

 

  During the setting process, if you do not need to calibrate the 

 

temperature, press the 

Enter

 key to skip the

 

/

 

 

or

 

/

 

 

 

option. 

  To exit the setup menu, press the 

Meas

 key. 

  The table below describes the relationship between the altitude 

 

and barometric pressure, make sure to set a compatible   

 

parameter before the calibration or measurement. 

Option 
Menu item 

 

Summary of Contents for DOscan10

Page 1: ...DOscan10 Pocket Dissolved Oxygen Tester USER MANUAL Bante Instruments Inc ...

Page 2: ...the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode ATC Indicates that the temperature compensation is enabled Prior to Use Filling the Electrolyte Solution 1 1 Take out the dissolved oxygen electrode from the carrying ...

Page 3: ...rement endpoint Enable Disable default Auto Power Off If enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Disable default Factory Reset If enabled all of the calibration values and current settings will be deleted or reset to the factory defaults the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode ...

Page 4: ...d the single point calibration in the setup menu we recommend that you perform a 100 saturation calibration in the air saturated water If the 2 points calibration is selected the zero oxygen solution needs to be used During the calibration and measurement the temperature sensor on electrode must be immersed in the sample solution completely the solution should keep 0 3 m s of minimum flow rate to ...

Page 5: ... hold the Cal key to enter the setup menu 2 Press the Enter key until the display shows or 3 Press the Cal key the tester enters the temperature calibration mode 4 Place the electrode into a solution with a known accurate temperature and wait for the measurement to stabilize 5 Press the Cal key to modify the temperature value 6 Press the Enter key to save and press the Meas key to return to the me...

Page 6: ... cap is contaminated or the electrolyte solution is depleted Screen shows Electrode is broken Replace the dissolved oxygen electrode 5 I DOscan10 1 Set the barometric pressure and salinity coefficient in the setup menu 2 Rinse the electrode with distilled water 3 Place the electrode into the sample solution and stir gently Wait for the measurement to stabilize and record the reading During the mea...

Page 7: ...ster is one year from the date of shipment Above warranty does not cover the electrode and electrolyte solution Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mo...

Reviews: