Preparation of Zero Oxygen Solution 

Dissolve 500 mg of the sodium sulfate (Na

2

SO

3

) reagent and a small 

amount of cobalt (II) chloride hexahydrate (CoCl

2

 

 

6H

2

O) in the 250 ml 

distilled water, mix the solution until reagent is completely dissolved. 
 

Tester Specifications 

 
Model 

DOscan10 

Dissolved Oxygen 

 

Range 

0.0 to 20.0 mg/L 

Resolution 

0.1 mg/L 

Accuracy 

± 0.5 mg/L 

% Saturation 

 

Range 

0.0 to 200.0% 

Resolution 

0.1% 

Accuracy 

± 2.0% 

Other Specifications 

 

Calibration Point 

1 or 2 points 

Temperature Compensation 

0 to 40° C (32 to 104° F), automatic 

Barometric Pressure Correction  450 to 850 mmHg, manual 

Salinity Correction 

0 to 35 g/L, manual 

Operating Temperature 

0 to 50° C (32 to 122° F) 

Storage Temperature 

0 to 60° C (32 to 140° F) 

Relative Humidity 

< 80% (non-condensing) 

IP Rating 

IP54 

Display 

LCD, 21 ×  21 mm (0.82 ×  0.82 in.) 

Power Requirements 

2 ×  1.5V AAA alkaline batteries 

Auto-Off 

8 minutes after last key pressed 

Dimensions 

175 (L) ×  40 (Ø )

 

mm (6.89 ×  1.57 in.) 

Weight 

100

 

g (3.5 oz.) 

 

Troubleshooting 

 
Fault 

Cause and Corrective Action 

Screen shows 

 

Dissolved oxygen electrode does not connect to the 
tester or measured value is out of range. 

Drifting erratic   
readings 

Check whether the membrane cap is contaminated 
or the electrolyte solution is depleted. 

Screen shows 

 

Electrode is broken. 
Replace the dissolved oxygen electrode. 

 
 

5   

I   

DOscan10 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

1.  Set the barometric pressure and salinity coefficient in the setup 
 

menu. 

2.  Rinse the electrode with distilled water.   
3.  Place the electrode into the sample solution and stir gently. Wait 
 

for the measurement to stabilize and record the reading. 

 

 

  During the measurement, if the display shows ---- indicating the 

 

measurement exceeds the range, remove the electrode from the 

 

sample solution immediately. 

  If the

 

 

option is enabled in the setup menu, the tester will 

 

automatically lock the measurement endpoint and show

 

HOLD 

 

icon. Press the 

Meas

 key to resume measuring. 

 

 

 

 

 

 

Electrode Maintenance 

  Rinse the electrode thoroughly with distilled water after use. 

  Do not touch the membrane and always keep it is clean and wet. 

  If you do not use the electrode for long periods, screw off the 

 

membrane cap, rinse the electrode anode, cathode, membrane 

 

cap with distilled water and blot dry. Install the electrode and 

 

store dry. 

 

 

 

 

 

 

 

Appendix 

 

Optional Accessories 

 
Order Code 

Description 

DO100 

Dissolved oxygen electrode, range: 0 to 20 mg/L 

DO-MEM 

Membrane cap, 2 PCS/set 

DO-ES 

Electrolyte solution, 30 ml 

 

Preparation of Air-Saturated Water 

Use an air-pump to blow air into distilled water at least 1 hour, while 
stirring the solution. 
 

Cathode

 

Anode

 

Summary of Contents for DOscan10

Page 1: ...DOscan10 Pocket Dissolved Oxygen Tester USER MANUAL Bante Instruments Inc ...

Page 2: ...the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode ATC Indicates that the temperature compensation is enabled Prior to Use Filling the Electrolyte Solution 1 1 Take out the dissolved oxygen electrode from the carrying ...

Page 3: ...rement endpoint Enable Disable default Auto Power Off If enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Disable default Factory Reset If enabled all of the calibration values and current settings will be deleted or reset to the factory defaults the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode ...

Page 4: ...d the single point calibration in the setup menu we recommend that you perform a 100 saturation calibration in the air saturated water If the 2 points calibration is selected the zero oxygen solution needs to be used During the calibration and measurement the temperature sensor on electrode must be immersed in the sample solution completely the solution should keep 0 3 m s of minimum flow rate to ...

Page 5: ... hold the Cal key to enter the setup menu 2 Press the Enter key until the display shows or 3 Press the Cal key the tester enters the temperature calibration mode 4 Place the electrode into a solution with a known accurate temperature and wait for the measurement to stabilize 5 Press the Cal key to modify the temperature value 6 Press the Enter key to save and press the Meas key to return to the me...

Page 6: ... cap is contaminated or the electrolyte solution is depleted Screen shows Electrode is broken Replace the dissolved oxygen electrode 5 I DOscan10 1 Set the barometric pressure and salinity coefficient in the setup menu 2 Rinse the electrode with distilled water 3 Place the electrode into the sample solution and stir gently Wait for the measurement to stabilize and record the reading During the mea...

Page 7: ...ster is one year from the date of shipment Above warranty does not cover the electrode and electrolyte solution Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mo...