6.2.1 Replacing fuses
Caution!
Repair work may only be carried out by authorised, quali
fi
ed personnel or by the
manufacturer. The manufacturer assumes no liability for unauthorised work on the
device!
Danger!
Always isolate from mains power before opening!
Risk of electric shock from live parts in the control box!
• Disconnect
the
fi
lter aggregate from mains power. Close the ball valves (ultrasonic
bath and
fi
lter aggregate).
• Remove the screws in the control box and open carefully.
• The fuse is inside the locking clip.
• Following inspection, replace only the defective fuse. The fuse rating is listed under
Technical Data, see Chapter 1.4.
• Reassemble the control box by following the above instructions in reverse.
• Reconnect the power cable. Re-open the ball valves.
6.3
Customer service
If service is required, please contact your specialist dealer or BANDELIN electronic to
order spare parts or before sending in the defective
fi
lter aggregate.
6.3.1 Repairs and service
If faults or defects have been discovered and cannot be recti
fi
ed, stop using the
fi
lter
aggregate immediately. If this occurs, please contact the supplier or the manufacturer:
BANDELIN electronic GmbH & Co. KG
Heinrichstraße
3-4
12207
Berlin
Repair service:
Email:
Tel.: +49 30 768 80-13
Fax: +49 30 768 802 00 13
Returns are subject to the General Conditions of Delivery and Payment of
BANDELIN electronic GmbH & Co. KG.
In addition, the
fi
lter aggregate must be cleaned and decontaminated as described in
the next chapter.