lter: If the pressure in the
fi
lter housing is too high, the pressure switch will
switch the pump off, and the red indicator light on the Off button will
illuminate.
Replace the
fi
lter cartridge.
Rinse the
fi
lter housing out thoroughly before replacing the
fi
lter cartridge.
After inserting a new
fi
lter cartridge and opening the three-way ball valve, check the
liquid level in the ultrasonic bath and
fi
ll if necessary.
This is particularly true for small cleaning tanks = small bath volume.
The
fi
lter aggregate can then be started again using the green On button - the red
indicator light will switch off.
6.2 Troubleshooting
SONOREX TECHNIK
fi
lter aggregates are robust constructions designed to offer
maximum reliability.
Nevertheless, the possibility of a malfunction due to a defective component can never
be ruled out entirely.
Caution!
Repairs must only be performed by authorised, quali
fi
ed personnel!
Unauthorised repairs will void the warranty!
Seq.
no.:
Malfunction
Possible cause(s)
1
Filter aggregate switches off, red
indicator light illuminates.
– Fine
fi
lter clogged
– Return
fl
ow throttled too far
– Return
fl
ow blocked, kinked or pinched
2
Filter aggregate switches off, red
indicator light does not illuminate.
– Mains power failure
– Defective fuse
3
Filter aggregate is running,
bubbles in inlet, little or no
fl
ow
volume in return
fl
ow.
– Bath drain closed
– Inlet blocked, improperly sealed, kinked
or pinched
– Pre-
fi
lter clogged
4
Filter aggregate running, hoses
fi
lled but no
fl
ow volume
– Pump blocked, magnetic coupling
slipping