22 / 38
23346-002 en/2023-03
Rinse sonication objects with clean water.
Do not leave sonication objects in the sonication fluid for too long.
This can damage the objects.
5.10 Emptying the oscillating tank
WARNING
Danger of electric shock
- Make sure that no liquid can get into the housing.
CAUTION
Hot sonication fluid and oscillating tank
There is a risk of burns when lifting the device to empty it.
- Allow the device to cool down before lifting it.
Contamination on the bottom of the oscillating tank reduces the ultrasonic output.
Empty and clean the oscillating tank if the sonication fluid is visibly contaminated.
Also, observe the specifications of the manufacturer of the agent regarding the service life
of the sonication fluid.
Replace used sonication fluid completely. Do not freshen the fluid by topping it up.
Procedure
1. Switch off the ultrasound. Switch off the heater (if applicable). If you need to move
the ultrasonic bath to empty it, disconnect the mains plug.
2. Open the 3-way ball valve and empty the oscillating tank.
3. Rinse the oscillating tank thoroughly.
4. Wipe the ultrasonic bath dry with a soft cloth.
5. If necessary, disinfect the ultrasonic bath with a suitable surface disinfectant.
Information
- Rinsing tanks with heater can also become hot.
Operation