![BANDELIN SONOREX PR 140 Instructions For Use Manual Download Page 22](http://html1.mh-extra.com/html/bandelin/sonorex-pr-140/sonorex-pr-140_instructions-for-use-manual_3692571022.webp)
22 / 27
51396c GB/2019-04
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
6.3
Error analysis
The ultrasonic bath is robustly constructed and designed for a high level of reliability.
Nevertheless, the possibility of a malfunction due to a defective component can never
be fully discounted.
The following overview of possible sources of error should serve as an aid for error
detection and elimination.
Ultrasonic bath oscillates weakly, unevenly, or noise is too loud:
- Has
fl
uid been properly degassed?
- Is it overloaded with objects to be treated?
- Uneven noises (wobbling)
Treat for 15 min.
Remove a few parts.
No error – slightly adjust the
fi
ll
level of the
fl
uid.
Heating system defective?
The ultrasonic bath can be readily operated
without heating.
Slight erosion visible on the tank bottom?
Natural wear.
Ultrasonic bath OK.
Any malfunctions are to be communicated in writing to the manufacturer.