4 Use
The volumetric glasware are directly sonicated, i.e. they are placed into the special
basket and submerged directly in the bath
fl
uid.
It is possible to soak the goods to be treated or the glass devices in a special storage
container without subsequent re-sorting by the tried-and-tested yesr system.
This loosens up stubborn deposits and shortens the sonication time.
4.1
Instructions for use
Instructions –
fi
lling
• Verify that the ball valve is closed.
• Ultrasound and heating must be turned off.
• Do not
fi
ll ultrasonic tank with hot water. Maximum
fi
lling temperature: 50°C.
• At least drinking-quality water must be used to
fi
ll the oscillating tank.
• Water without additives is not suited for sonication. BANDELIN recommends the
TICKOPUR or STAMMOPUR preparations.
• Do not use distilled or deionised water without additives.
• The
fi
ll level must always be at least at the lower
fi
lling level mark.
A low
fi
ll level will damage the ultrasonic bath!
• Thoroughly rinse parts that have come in contact with aggressive chemicals (acids,
chloride ions), before sonication.
• When using preparations, the safety instructions included in the respective product
lea
fl
ets must be fundamentally adhered to.
• Do not use any combustible
fl
uids (e.g. benzine, solvents) or chemicals that contain
chloride ions or that separate (some disinfectants, household cleaners, and dish
detergents), for sonication in the stainless steel tank.
• Replace used bath
fl
uids, do not refresh by adding
fl
uids.