42
7.
Lift one end of the unit up a minimum of 5” and block in position.
8.
Unscrew the bolts (one per leg) that hold the legs in the shipping position.
9.
Slide the leg out of the stand until the holes line up at the next position.
10.
Bolt the leg in position with items 2, 3 and 4 (two bolts per leg).
11.
Repeat steps 8 and 9 for the other leg and remove blocks.
12.
Repeat steps 6 through 10 for the other end of the unit.
13.
Plug open holes using item 5 (one per leg).
To raise the unit to the 38 5/8” work-surface height (adjustable foot option)
To raise the unit to the 36” work-surface height (caster option)
14.
Lift one end of the cabinet up a minimum of 7” and block in position.
15.
Unscrew the bolts (two per leg) that hold the legs in the current work-surface-height position.
16.
Slide the leg out of the stand until the holes line up at the next position.
17.
Bolt the leg in position with items 2, 3 and 4 (two bolts per leg).
18.
Repeat steps 15 and 16 for the other leg and remove blocks.
19.
Repeat steps 13 through 17 for the other end of the unit.
20.
Plug open holes using item 5 (two per leg
adjustable foot option
)
(one per leg
caster option
)
Turn in opposing directions for fine adjustment to aid in leveling of cabinet.