23
Data Trend Log
The Data Trend Log screen can be accessed from the LOG button on screen 6.
Screen 7 shows the Trend Log display. The earliest recorded data will be displayed. The window can
be scrolled through on the control screen. To view previous data, the USB is to be removed and
connected to a computer. The USB can capture approximately 6 months of data.
Delete: Press delete button and the data will disappear from the system.
USB Save: Press USB Save button and the data log will downloaded to a USB.
Figure 32: Screen 7
–
Trend Log
Engineer Mode Screen
To set the control parameters, enter the Engineer Mode screen. Entry to the Engineer Mode screen
is available when the time shows 29s-30s or 59s-00s. Double click on the corner of the top right of
the atmosphere control screen, and then you will enter the engineer screen. Press the value, and a
pop-up will appea
r for the new setting to be selected, then press “save” button and “back” button,
and you will return to the running mode.
Figure 33: Screen 8
–
Engineer Mode Screen