July 2021
18
FAULT
EVENT LOG INDEX ERR,
PID LOCK VALUE ERR,
Firmware Revisions Incompatible
This is a generic fault
event message.
Arguments indicating
the type of fault will be
appended to the
message. Please
consult engineering for
assistance deciphering
the appended fault
code.
User Action Events
Event Name
Event
Code
Event Code Meaning
Description
ALARM MUTE
n/a
The Audible Alarm has
been muted by the user.
UV DECON ON
n/a
The user has enabled
the UV
Decontamination.
UV DECON OFF
n/a
The user has disabled
the UV
Decontamination.
TEMP SET
<value>
The new set point value.
The “Air Temperature
Set Point” value has
changed.
HI TEMP SET
<value>
The new HI alarm set point.
The “Air Temperature
Alarm High” set point
value has changed.
LO TEMP SET
<value>
The new LO alarm set point.
The “Air Temperature
Alarm Low” set point
value has changed.
TEMP OFFSET
<value>
The new OFFSET value.
The “Air Temperature
Offset” value is
changed.
TEMP SLOPE
<value>
The new SLOPE value.
The “Air Temperature
Slope” value is changed
RH SET
<value>
The new set point value.
The “RH Set Point” value
has changed.