Instrument Overview
12/3/2009 | 71-022 V6
D30R
16
24)
E-stop lights connector
Port for external large lights that fit on top of unit for added safety.
Initial power-up and check-out
Note:
Each Baker Instrument Company, an SKF Group Company Tester incorporates a
supply ground detection circuit. This circuit assures a positive grounding of the tester. If the
instrument is not properly ground, the OPEN GRN indication will light and testing cannot
proceed. Check the supply to the tester (broken ground, bad extension cord, excessive ground
to neutral voltage) and assure that a low impedance ground is provided to the unit.
1)
Check that the On/Off switch is in the Off position.
2)
Connect the tester power cord to a 110-120 VAC outlet (or 220-240 VAC outlets if
appropriate). The tester will work on either 60 Hz or 50 Hz.
3)
Set the Output control to Min (fully counterclockwise).
4)
Turn the tester On/Off switch to On. Allow a brief period for CRT warm-up. The
following message should appear on the screen.
A self-test performed on the tester for all internal memory modules when the tester is
powered up. Any failures will be noted on the CRT display.
Using the footswitch
The Digital tester can be equipped with a footswitch that allows hands free operation of the
instrument. For example it is used to help eliminate the display effect of rotor loading, by
allowing the operator to spin the rotor of the equipment under test.
Summary of Contents for D30R
Page 6: ...Table of Contents 12 3 2009 71 022 V6 D30R 6...
Page 12: ...Safety precautions CE compliance 12 3 2009 71 022 V6 D30R 12...
Page 24: ...Coil Resistance Testing 12 3 2009 71 022 V6 D30R 24...
Page 28: ...Principles of High Voltage DC Testing 12 3 2009 71 022 V6 D30R 28...
Page 38: ...High Voltage DC Testing 12 3 2009 71 022 V6 D30R 38...
Page 50: ...Surge Testing 12 3 2009 71 022 V6 D30R 50...
Page 70: ...Data Recording Retrieving 12 3 2009 71 022 V6 D30R 70...