38
7.2 Standard software for
O
2
analysis
Measurement range examples
• 0 to 10; 100; 1,000; or 10,000 ppm O
2
• 0% to 1%; 10%; 21%; 25%; or 100% O
2
Relays
• Four process alarms
• One fault alarm
• Two automatic calibration contacts
• Two calibration alarms
7.3 Hydrogen-cooled
generator software
Three ranges
• 0% to 100% H
2
in air
• 0% to 100% H
2
in CO
2
• 0% to 100% air in CO
2
Relays
• Two process alarms
• One fault alarm
• One normal alarm
7.4 Physical specifications
Dimensions (w x h x d)
9 in x 10 in x 9 in. (229 mm x 254 mm x 229 mm)
Weight
15 lb (6.8 kg)
Environmental/certifications
•
Weatherproof:
type 4X enclosure: IP66
•
Explosion-proof:
type 7 enclosure: FM/CSA class I,
division 1, group B, C and D
•
Flameproof:
Intertek ITS12ATEX17703X II 2 G Ex d IIC T6 Gb
IECEx ITS 12.0058x
European compliance
Complies with EMC directive 2014/30/EU, 2014/35/EU LVD
(installation category II, pollution degree 2)
Conduit entry
Six, 3/4” NPTF conduit ports
Mounting holes
Four, 3/8” (10 mm) holes
7.5 Order information
To ensure that the appropriate XDP is supplied, a full and
accurate model number must be specified at the time
of purchase. This model number should be constructed
as follows:
XDP -
B
C
D
E
B: Package
2 = Explosion-proof/weatherproof
C: Power
2 = 100-240 VAC
5 = 24 VDC
D: Keypad
1 = Magnetic
E: Software
1 = Standard
2 = Hydrogen cooled generator*
*
DO NOT
use this manual with software option 2!
For example, an XDP in an explosion-proof enclosure with
standard software, a magnetic keypad, and a 100-240 VAC
input voltage requirement would be specified as:
XDP-2211
.
Summary of Contents for Panametrics XDP
Page 1: ...Gas analysis XDP User s manual 910 204 Rev E...
Page 2: ...ii...
Page 4: ...no content intended for this page proceed to next page iv...
Page 8: ...viii no content intended for this page...
Page 12: ...no content intended for this page proceed to next page 2...
Page 28: ...no content intended for this page proceed to next page 18...
Page 36: ...26 no content intended for this page proceed to next page...
Page 42: ...32 no content intended for this page proceed to next page...
Page 58: ...48 no content intended for this page proceed to next page...