3.4 Powering up the system
Check the wiring connections and close both the XDP and
transmitter enclosures before applying power. Energize the
external disconnect device to power up the XDP. Allow the
transmitter to warm up for at least thirty minutes, before
taking measurements.
WARNING!
The XDP requires an external power
disconnect device such as a switch or
circuit breaker. The disconnect device must
be marked as such, clearly visible, directly
accessible, and located within 1.8 m (6 ft) of
the unit.
3.4.1 The enclosure window
The window at the top of the XDP enclosure includes the
components shown in
Figure 9
below.
XDP
Display window
ENTER key
RIGHT arrow
DOWN arrow
UP arrow
LEFT arrow
ESC key
Figure 9: The enclosure window
The XDP is factory-configured to display
O2 ppm/pcnt with
a resolution of 0.01
as the default. However, changing the
gas label, the gas units, and the display resolution are easily
accomplished. See
, for
programming instructions.
IMPORTANT
The XDP magnetic keypad permits programming of the
instrument through the glass faceplate without removing
the cover. Thus, all programming procedures may be
performed while the unit is installed in a hazardous area.
3.4.2 Initial screen displays
Immediately upon power up, the digital display shows the
following sequence of informational screens, as the meter
performs its internal systems checks.
This is a typical example
of the first initialization
screen.
This is a typical example
of the second initialization
screen.
This is a typical example
of the final initialization
screen.
If the meter passes all of its internal system checks, the
following initial data screen appears:
Notice that the absence
of a valid signal has
triggered the low low
alarm (the LL symbol in the
lower right corner).
NOTE:
The padlock icon at the upper right hand corner of
the above display indicates that the
user program
is locked
with password protection.
3.4.3 Starting the sample gas
Start the sample gas flow to the transmitter and allow a few
minutes for the system to reach equilibrium. The following
data screen should appear if the oxygen level is above the
low alarm setpoint and below the high alarm setpoint.
This is a typical display.
The actual oxygen
concentration of your
sample gas will be shown
here.
NOTE:
Allow approximately 30 minutes for the sensor to
reach temperature equilibrium, to ensure accurate readings
Notice that the display shows the
slot number
, the
gas label
,
and the
current time
on the top line. The current
data value
along with the
units
are displayed across the bottom of
the screen.
NOTE:
Slot 0 is the main circuit board, slot 1 holds the XDP
circuit board, and slots 2-4 are normally not used.
Summary of Contents for Panametrics XDP
Page 1: ...Gas analysis XDP User s manual 910 204 Rev E...
Page 2: ...ii...
Page 4: ...no content intended for this page proceed to next page iv...
Page 8: ...viii no content intended for this page...
Page 12: ...no content intended for this page proceed to next page 2...
Page 28: ...no content intended for this page proceed to next page 18...
Page 36: ...26 no content intended for this page proceed to next page...
Page 42: ...32 no content intended for this page proceed to next page...
Page 58: ...48 no content intended for this page proceed to next page...