3.3.7 Trimming the output (cont.)
Example:
1. Trim is reset, then Trim Zero is selected. The connected output device reports 3.977 mA.
2. The operator enters “3.977” as the Zero Trim value.
3. Trim Span is selected. The connected output device reports 19.985 mA.
4. The operator enters “19.985” as the Span Trim value.
5. The
dew.IQ
adjusts the output accordingly to align the output with the readings by the connected output device or a
DVM.
6. Using the Test Menu, the operator verifies that a test value of 0% now reads 4.000 mA at the connected output device,
and a test value of 100% now reads 20.000 mA.
3.4 Setting up the measurement alarms
The
dew.IQ
has with two programmable high/low alarms and one fault alarm relay. Use the instructions in this section to set
up these alarms.
3.4.1 Selecting an alarm output
To set up the alarms, on the Main Menu choose Alarm and press
Enter
. Then, from the Alarm
Menu choose Select and press
Enter
. A screen similar to the following appears:
Use the arrow keys to select the output (A or B) to be set up and press
Enter
. The display
returns to the Alarm Menu.
46
Summary of Contents for Panametrics dew.IQ
Page 1: ...dew IQ Moisture analyzer User s manual 910 295 Rev C...
Page 2: ...ii...
Page 3: ...dew iQ Moisture analyzer User s manual 910 295 Rev C October 2017 i...
Page 4: ...no content intended for this page ii...
Page 63: ...no content intended for this page 56...
Page 101: ...no content intended for this page 94...
Page 103: ...Figure 39 Wall Mount Outline and Installation ref dwg 712 1823 1 of 3 96...
Page 104: ...Figure 40 Wall Mount Outline Installation ref dwg 712 1823 2 of 3 97...
Page 105: ...Figure 41 Wall Mount Outline Installation ref dwg 712 1823 3 of 3 98...
Page 106: ...Figure 42 Rack Mount Outline Installation ref dwg 712 1824 99...
Page 107: ...Figure 43 Panel Mount Outline Installation ref dwg 712 1825 100...
Page 108: ...Figure 44 Bench Top Outline Installation ref dwg 712 1826 101...
Page 109: ...Figure 45 Interconnection Diagram ref dwg 702 1381 102...
Page 111: ...FIgure 46 Main Menu Map Using M Series Probe 104...
Page 112: ...FIgure 47 Main Menu Map Using IQ probe 105...
Page 113: ...no content intended for this page 106...