3-30
CL 5
Creating and Erasing Custom Setup Files
Setting Up the CL 5
alarm is violated. Note that the inputted alarm value
will correspond to a thickness except when the VELO-
CITY measurement option is installed and activated.
In this case the alarm settings will represent velocity
values. To input a minimum or maximum alarm:
Step 1:
Press to access the Configuration display
(shown in Figure 3-3).
Step 2:
Press or to select
MINIMUM ALARM
or
MAXIMUM ALARM
Step 3:
When the desired parameter is highlighted,
press
to activate it.
Step 4:
Press or to change the selected posi-
tion in the alarm thickness and or to adjust the
value of the selected position. To switch the alarm
setting to OFF from any other value, simultaneously
press an .
Step 5:
Press when the alarm thickness is set to
the desired value.
3.5 Creating and Erasing
Custom Setup Files
After loading a Standard Setup and performing the
necessary adjustments to optimize performance for
a given application, the instrument settings can be
stored and named as a Custom Setup file. The instru-
ment is capable of storing up to 5 Custom Setup files,
which can be erased at any time
(Figure 3-5).
Custom Setups can be created from any CL 5 key-
pad. Custom Setups created using a base-model
instrument include the following settings:
• User-assigned Custom Setup file name
• VELOCITY
• MINIMUM ALARM
• MAXIMUM ALARM
Summary of Contents for 022-508-909
Page 1: ...Krautkr mer CL 5 Operating Manual 021 002 296 rev 04...
Page 2: ...CL 5 1 2 General Information 1...
Page 10: ...CL 5 2 10 Understanding the Keypad Menu System and Displays 2...
Page 20: ...CL 5 3 20 Setting Up the CL 5 3...
Page 36: ...CL 5 4 36 Measuring Thickness 4...
Page 49: ...CL 5 5 49 Using the Optional Data Recorder 5...
Page 57: ...CL 5 6 57 I O Features 6...
Page 59: ...CL 5 6 59 Transmitting Thickness Data Through an External Device I O Features...
Page 69: ...CL 5 7 69 Specifications 7...
Page 77: ...CL 5 8 77 Maintenance 8...
Page 79: ...CL 5 9 79 Appendix 9...
Page 84: ...CL 5 10 84 Index 10...