CL 5
3-23
The following instrument settings are automatically
adjusted when the Setup file is activated:
View Mode
Gain
Measured Value display resolution
Minimum Alarm Setting
Maximum Alarm Setting
Zero Offset – for contact probes only
Velocity
Nominal Thickness
Note:
While Custom Setups can be created in any
instrument, the parameter settings, which may be
modified and then saved in a Custom Setup, depend
on the instrument’s configuration. In a base model
instrument, only the setup name, material velocity
value (determined via calibration), alarm thickness
values, and UT settings related to the probe setup
can be modified and stored in a custom setup.
Setting Up the CL 5
Configuring the Instrument‘s Display
Summary of Contents for 022-508-909
Page 1: ...Krautkr mer CL 5 Operating Manual 021 002 296 rev 04...
Page 2: ...CL 5 1 2 General Information 1...
Page 10: ...CL 5 2 10 Understanding the Keypad Menu System and Displays 2...
Page 20: ...CL 5 3 20 Setting Up the CL 5 3...
Page 36: ...CL 5 4 36 Measuring Thickness 4...
Page 49: ...CL 5 5 49 Using the Optional Data Recorder 5...
Page 57: ...CL 5 6 57 I O Features 6...
Page 59: ...CL 5 6 59 Transmitting Thickness Data Through an External Device I O Features...
Page 69: ...CL 5 7 69 Specifications 7...
Page 77: ...CL 5 8 77 Maintenance 8...
Page 79: ...CL 5 9 79 Appendix 9...
Page 84: ...CL 5 10 84 Index 10...