Operating Manual BaehrTec S2000
English– V101 – Status 07/2015
Page 49/52
Guidelines and MANUFACTURER statement
– ELECTROMAGNETIC EMISSION
The model BaehrTec S2000 is intended for use in the ELECTROMAGNETIC ENVIRONMENTS specified below. The customer or
user
N11)
of the model BaehrTec S2000 should ensure that it is used in such an environment.
Emission measurements
Compliance
ELECTROMAGNETIC ENVIRONMENT – Guidelines
HF emissions
according to CISPR 11
Group 1
The model BaehrTec S2000 uses HF energy exclusively in its
internal FUNCTION. That means that its HF emission is very low
and it is unlikely that adjacent electronic devices will suffer any
interference.
The model BaehrTec S2000 is intended for use in all facilities
including living areas and also that are connected to a PUBLIC
POWER SUPPLY that include buildings that are used for
residential purposes.
HF emissions
according to CISPR 11
Class B
Harmonics
under IEC 61000-3-2
Class A
Voltage fluctuations/
flicker under IEC 61000-3-3
Fulfilled
N11)
National footnote: Here, user is meant in the sense of a "RESPONSIBLE ORGANISATION".
Guidelines and MANUFACTURER STATEMENT
–
ELECTROMAGNETIC EMISSION
The model BaehrTec S2000 is intended for use in the ELECTROMAGNETIC ENVIRONMENTS specified below. The
customer or user
N13)
of the model BaehrTec S2000 should ensure that it is used in such an environment.
STABILITY
TESTS
IEC 60601-
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT –
GUIDELINES
ELECTROSTATIC
DISCHARGE (ESD)
under IEC 61000-4-2
± 6 kV
Contact discharge
± 8 kV
Air discharge
± 6 kV
Contact discharge
± 8 kV
Air discharge
Floorings should be
made of wood or
concrete or covered with
tiles. If the floor is
covered with synthetic
material, the relative
humidity must be at least
30 %.
Electrical
electric
disturbance variables/bursts
according to IEC 61000-4-4
± 2 kV
for power supply lines
± 1 kV
for input and
output lines
± 2 kV
For power supply lines
± 1 kV
For input and
output lines
The quality of the supply
voltage should be to the
standard of a typical
commercial or hospital
environment.
Surge voltages
under IEC 61000-4-5
± 1 kV
Voltage phase
conductor -
phase conductor
± 2 kV
Voltage phase
conductor -
earth
± 1 kV
Voltage phase conductor -
phase conductor
± 2 kV
Voltage phase conductor -
earth
The quality of the supply
voltage should be to the
standard of a typical
commercial or hospital
environment.
Voltage dips,
Short interruptions
and fluctuations in the
supply voltage
according to IEC 61000-4-11
< 5 %
U
T
for 1/2 period
(> 95% dip)
40 %
U
T
for 5 periods
(60 % dip)
70 %
U
T
for 25 periods
(30 % dip)
< 5 %
U
T
for 5 s
(> 95% dip)
< 5 %
U
T
for 1/2 period
(> 95% dip)
40 %
U
T
for 5 periods
(60 % dip)
70 %
U
T
for 25 periods
(30 % dip)
< 5 %
U
T
for 5 s
(> 95% dip)
The quality of the supply
voltage should be to the
standard of a typical
commercial or hospital
environment.
We recommend powering
the model BaehrTec
S2000 with a interruption-
free power supply or a
battery, should the user
of the model BaehrTec
S2000 FUNCTION also
require continued
FUNCTION also when
interruptions in the power
supply occur.
Magnetic field at a supply
frequency of
(50/60 Hz)
under IEC 61000-4-8
3 A/m
3 A/m
The magnetic fields of
both power frequencies
should be to the standard
of a typical commercial or
hospital environment.
N13)
National footnote: Here, user is meant in the sense of a "RESPONSIBLE ORGANISATION".